24.11.2012 Views

Extraction Technologies For Medicinal And Aromatic Plants - Unido

Extraction Technologies For Medicinal And Aromatic Plants - Unido

Extraction Technologies For Medicinal And Aromatic Plants - Unido

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

EXTRACTION TECHNOLOGIES FOR MEDICINAL AND AROMATIC PLANTS<br />

ii) 2–3 ml/min for a semi preparative column with 1.6 mm i.d.<br />

PTFE tubing (800–1000 rpm) (up to 500 mg sample load);<br />

iii) 1 ml/min for an analytical column with 0.85–1.0 mm i.d.<br />

PTFE tubing (1000–1200 rpm) (up to 50 mg sample load).<br />

• These fl ow rates should be modifi ed according to the settling<br />

time of the two-phase solvent system as well as other factors.<br />

When the settling time is around 20 s and the K value<br />

of the analyte is small, a lower fl ow rate is recommended.<br />

13.5.1.7 Revolution Speed<br />

The optimum revolution speed (revolution and planetary rotation<br />

speeds are always same) for the commercial HSCCC-CPC instrument for<br />

preparative separation ranges between 600 and 1400 rpm.<br />

• Use of a lower speed will reduce the volume of the stationary<br />

phase retained in the column, leading to lower peak resolution.<br />

• On the other hand, the higher speeds may produce excessive<br />

sample band broadening by violent pulsation of the column<br />

because of elevated pressure.<br />

13.5.1.8 Filling the Column with the Stationary Phase<br />

• In each separation, the column is fi rst entirely fi lled with the<br />

stationary phase.<br />

• Before introducing the stationary phase, the column may be<br />

fl ushed with a column volume of a solvent miscible with the<br />

two phases used in the previous run (e.g. ethanol or methanol)<br />

to wash out materials remaining in the column.<br />

• This will also ensure a stable, clean baseline before the solvent<br />

front emerges.<br />

• Avoid trapping the air in the column, especially in a preparative<br />

column. This can be easily tested as:<br />

i) If no air is present in the column, the fl ow from the column<br />

outlet is ceased shortly after stopping the pumping.<br />

ii) If the solvent keeps fl owing from the outlet for more than<br />

several seconds, the air trapped in the column should<br />

be eliminated by resuming the pumping of the stationary<br />

phase under low speed column rotation (100–200 rpm)<br />

in a tail to head elution mode to accelerate air movement<br />

toward the outlet of the column.<br />

13.5.1.9 Sample Loading<br />

There are two ways to load samples; both are satisfactory for<br />

HSCCC-CPC separations:<br />

• In the fi rst method, the column is entirely fi lled with stationary<br />

phase and this is immediately followed by sample injection.<br />

225

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!