Model 735 Intelligent Display User's Manual - Paroscientific, Inc.
Model 735 Intelligent Display User's Manual - Paroscientific, Inc.
Model 735 Intelligent Display User's Manual - Paroscientific, Inc.
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
CAUTION PF must be set to the full-scale pressure valueof the Digiquartz® transducer. Failure to do so will resultin improper formatting of pressure measurement values.The example below is for a 45 psi transducer.Typical set command:Typical set response:*0100EZ*0100PF=45*0001PF=45.00000• Set the transducer overpressure limit valuePL is typically set to 1.2 times the full-scale pressure value ofthe Digiquartz® transducer. Refer to the transducerSpecification Control Drawing for specific overpressurelimits. PL is always set and reported in psi, regardless of thepressure unit setting.The example shown below is for a 45 psi transducer (45 x 1.2= 54).Typical set command:Typical set response:*0100EZ*0100PL=54*0001PL=54.00000• Set the transducer type valueSet PO according to the Digiquartz® pressure transducertype (absolute, gauge, or differential). The table below showsthe integer values used with PO to identify the transducertype:QUICK START 3-4