PSSet or read the temperature measurement interval used by the P7command.Action: Controls how often temperature measurements are takenduring a P7 pressure measurement sequence.If PS=0, an initial temperature measurement is taken, andall subsequent pressure measurements are compensatedusing that value.If PS=1, a temperature measurement is taken before eachpressure measurement (same as P4).If PS=n and n>1, an initial temperature measurement istaken, and subsequent temperature measurements aretaken after every n pressure measurements.Range: 1 to 65535Typical command: *0100EW*0100PS=4Typical response: *0001PS=4Note: When setting virtually all parameter values, you mustprecede the command with an EW (Enable Write)command. Please refer to paragraph 5.7.2.1 for moreinformation.Q2Continuously sample and send temperature period measurementvalues in units of microseconds.Action: Measure temperature period, send temperature periodvalue, and repeat until commanded to stop.Typical command: *0100Q2Typical response: *00015.1234567 (Value: 5.1234567)*00015.1234568 (Value: 5.1234568)*00015.1234567 … (Value: 5.1234567)Note: The Q2 command will temporarily prevent the front paneldisplay from being updated.REMOTE OPERATION 5-13