5.7.1.3 Continuous Measurement CommandsThe following commands are used to initiate and control continuousmeasurements. Continuous measurement commands repeatedly takemeasurement samples and return measurement values until commanded tostop. Continuous measurements are cancelled by sending any validcommand.P2Continuously sample and send pressure period measurementvalues in units of microseconds.Action: Measure pressure period, send pressure period value, andrepeat until commanded to stop.Typical command: *0100P2Typical response: *000128.123456 (Value: 28.123456)*000128.123457 (Value: 28.123457)*000128.123456 … (Value: 28.123456)Note: The P2 command will temporarily prevent the front paneldisplay from being updated.P4Continuously sample and send pressure measurement values inselected engineering units.Action: Measure temperature period, measure pressure period,calculate temperature-compensated pressure, sendpressure value, and repeat until commanded to stop.Typical command: *0100P4Typical response: *000114.71234 (Value: 14.71234)*000114.71235 (Value: 14.71235)*000114.71234 … (Value: 14.71234)REMOTE OPERATION 5-11