Checking the meter 7did you perform the control test according tothe user’s <strong>Manual</strong>?did you wait for the application prompt to bedisplayed before applying control solution?did you apply the control solution immediatelyafter opening the applicator?did you enter the correct color?did you perform the test within the correcttemperature range (10–40 °C)?yesyesyesyesyesHas the expiry date for the control solutionbeen reached (see label on the bottom of thevial next to the symbol )?did you read the section “Storage and stabilityof control solutions” in the package insertfor the control solution?did you observe the proper storage conditionsfor the meter, test cassette and controlsolutions (see Chapter 15 and the test cassetteand control solution package inserts)?noyesyesIf your meter has been dropped, that can also lead tocontrol test results which are not within the concentrationrange displayed and for which Control test not OKis displayed. In this case, you should also call<strong>Accu</strong>-<strong>Chek</strong> Customer Care.187