28.01.2013 Views

UNIVERSIT . . AT BONN Physikalisches Institut - Prof. Dr. Norbert ...

UNIVERSIT . . AT BONN Physikalisches Institut - Prof. Dr. Norbert ...

UNIVERSIT . . AT BONN Physikalisches Institut - Prof. Dr. Norbert ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

7.1. Dynamic range 95<br />

7.1.1 Counter dynamic range<br />

Fig. 7.1 showed the characteristic dynamic range of the CIX 0.2 system. In order to<br />

generalize this result, it has to be analyzed with respect to the theoretically expected<br />

maximum count rate, the impact of the detector bias on the maximum rate and potential<br />

limitations of the counter design.<br />

Maximum count rate<br />

The previous section stated that at 10 keV threshold and maximum feedback settings<br />

(IF b = 91 nA), the system’s maximum count rate lies at 3.3 Mcps. This value was measured<br />

at a tube voltage of 90 keV, which, according to section 6.4.3, corresponds to an average<br />

measured photon energy of approximately 29 keV. A direct crosscheck of this result with<br />

the internal charge injection circuits of CIX 0.2 is not possible because the average pulse<br />

size of 29 keV lies below the minimum input charge of 1.1 fC (32 keV) that the test circuits<br />

can deliver (see section 4.2.1). Nevertheless, the expected maximum count rate at 29 keV<br />

pulse size can be extrapolated from the maximum count rate measurements presented in<br />

section 4.3.1. Fig. 4.8 showed the maximum count rate for different pulse sizes at two<br />

different threshold settings. Extrapolating the curve at 10 keV threshold setting to a<br />

pulse size of 29 keV gives a maximum pulser rate of (8.8 ± 0.2) MHz. However, as the<br />

time interval between individual photons in the X-ray beam is not constant but rather<br />

Poisson-distributed, the value of 8.8 MHz is too large. Section 3.3.1 stated that the count<br />

rate behavior of a paralyzable counter, if exposed to monoenergetic Poisson-distributed<br />

pulses, is given by (3.2). Replacing the incident photon spectrum by a Poisson-distributed<br />

series of monoenergetic photons at exactly the spectrum’s average energy yields:<br />

N Max<br />

Meas = (8.8 ± 0.2) MHz · e −1 = (3.24 ± 0.08) Mcps (7.1)<br />

As the X-ray measurements yielded a maximum rate of (3.3 ± 0.13) Mcps, this shows that<br />

the electrical test and the X-ray measurements are in excellent agreement.<br />

Bias-dependence<br />

Besides the discriminator setting and the spectral distribution of the input pulses, the<br />

maximum count rate in a direct converting semiconductor sensor is also influenced by the<br />

applied bias. Hence, the maximum count rate of different CIX 0.2 modules equipped with<br />

CdTe and CdZnTe sensors was evaluated at various bias settings. Note that the very low<br />

X-ray absorption probability of Si does not allow a similar measurement for the Si module<br />

as the ASIC’s maximum count rate can not be reached with the available X-ray tube.<br />

Therefore Fig. 7.2 only contains the results for the Cd-based sensor materials.<br />

The measurements show that at low bias values the maximum count rate increases with<br />

the bias voltage in all sensor samples. This is a direct consequence of the higher electric<br />

field inside the sensor, which reduces the transit time of the generated charge carriers and<br />

therefore the pulse width and the comparator dead time. In the case of the 1 mm thick<br />

CdTe sample, the maximum count rates are comparable to those of CZT04. However, unlike<br />

CdZnTe the optimum bias value is not equal to the highest safely achievable voltage<br />

across the detector. Instead, a bias of 300 V offers the highest count rate capability. The<br />

reason for this is the photon flux-dependent current, which manifests itself in the CdTe<br />

samples and causes a baseline shift at the preamplifier output (see sections 6.1.3).<br />

In addition, Fig. 7.2 also illustrates that the performance among the CdZnTe samples<br />

varies greatly. The previous section showed that the maximum count rate of the CZT04

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!