28.01.2013 Views

UNIVERSIT . . AT BONN Physikalisches Institut - Prof. Dr. Norbert ...

UNIVERSIT . . AT BONN Physikalisches Institut - Prof. Dr. Norbert ...

UNIVERSIT . . AT BONN Physikalisches Institut - Prof. Dr. Norbert ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

6.2. Temporal response 73<br />

amplitude changes.<br />

Furthermore, a distinctive rectangular pattern is visible in the integrator image. This<br />

is the result of the stitching of subsequent measurements because during irradiation the<br />

decrease in signal amplitude varies from pixel to pixel. Tiling the individual measurements<br />

to form the large 9 x 6 view then creates the repetitive pattern. Hence, the first two images<br />

illustrate the need for a good long-term stability of the sensor signals.<br />

The effect of a slow response to a varying input signal is demonstrated in the third image<br />

(see Fig. 6.7(c)). This image shows the fluctuations in the current signals on a short<br />

(1 ms) timescale using the dead time free readout of CIX. At each x-y stage position<br />

10 individual snap-shots of that particular image section were taken and the standard<br />

deviation of these 10 measurements was calculated. Although the long-term decrease in<br />

current amplitude is lost in this measurement, distortions at the edges of the cogwheel are<br />

visible. This is because during the imaging procedure the middle section of the cogwheel<br />

was scanned going from left to right. Like this the remainders of high current signals from<br />

outside the object are superimposed over the low signal section of the cogwheel. Similarly,<br />

when moving out of the brass absorber the current signals only slowly return to the high<br />

amplitudes.<br />

6.2.1 Long-term response<br />

The afore mentioned long-term change in the detector response of CdTe and CdZnTe was<br />

studied in detail for different bias and flux conditions. In these measurements the sensors<br />

were irradiated with a constant beam intensity for five minutes. Following this irradiation<br />

period, the sensor signals were recorded for another five minutes without irradiation. The<br />

irradiation interval of five minutes was chosen as it reflects the typical scan durations<br />

when imaging a larger object with CIX 0.2. In order to evaluate the ASIC’s performance<br />

under such measurement conditions, identical measurements were performed on a 300 µm<br />

thick Si sensor. In this context they will serve as a reference since Si is not limited by<br />

comparable temporal instabilities. The focus of this analysis is placed on the integrator<br />

because of its inherent sensitivity to slow changes in the signal amplitude. In contrast to<br />

this, the CIX counter is only influenced indirectly via a preamplifier baseline shift.<br />

Integrator long-term response - Pulse shapes and mathematical description<br />

Fig. 6.8 shows the time dependence of the average pixel current for a Si, a CdZnTe and<br />

a CdTe sensor at a tube current of 30 µA at 6.5 cm focal spot to detector distance and<br />

at high bias settings. As expected, the Si sensors exhibit no significant temporal changes<br />

during or after an irradiation period. This is not true for the CdTe and CdZnTe detectors<br />

where a significant time-dependence of the signals is observed. Note that the curves in<br />

Fig. 6.8 are representative examples, which were confirmed on other samples of the three<br />

sensor materials.<br />

Starting with the CdZnTe detectors, the current under X-ray irradiation IOn can be described<br />

by a combination of an exponential function with a time constant τOn, specifying<br />

the initial amplitude increase, and a linear function, which parameterizes the subsequent<br />

rise. The observed linear runaway of the detector currents under irradiation has recently<br />

been confirmed in a separate study on CdZnTe samples with high µhτh products at high<br />

bias [53]. At the end of the 5 min irradiation period, the current IOff quickly drops down<br />

to a few percent of the current under irradiation and then slowly returns back to the baseline<br />

following an exponential tangent with the time constant τ long<br />

Off . Even on a timescale<br />

of one hour no current undershoot after tube shut-off is observed. The mathematical de-

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!