Calibration Procedure Diagnostics<br />
6. Adjust R81 potentiometer and verify the linearity accordingly. See figure below.<br />
Static Calibration Setup<br />
Digital Mercury Manometer<br />
Bulb<br />
150 mmHg<br />
FIGURE 6-13<br />
Bulb<br />
Valve<br />
Manually<br />
Squeeze<br />
Bulb<br />
T - Fitting<br />
NIBP Module<br />
693cc<br />
Test Chamber<br />
P/N 0138-00-0001-01<br />
Specification: = 150mmHg +/- 0.1%<br />
Previous Menu<br />
Overpressure Voltage Setpoint Static Pressure Calibration<br />
Static Pressure Calibration<br />
Motor Pump Test Transducer: 150.0 mmHg<br />
Leak Test<br />
Dump Test<br />
Bleed Rate Test<br />
Overpressure Test<br />
Abort Test<br />
7. Reassemble unit and verify the following static pressure points to ensure proper<br />
calibration.<br />
Specification: 30 mmHg, 100 mmHg, 150 mmHg, 190 mmHg and 250<br />
mmHg. +/-3 mmHg.<br />
<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Service Manual 0070-10-0441 6 - 11<br />
VR 81<br />
Select to return to previous menu.<br />
NIBP Test<br />
<strong>Passport</strong> 2<br />
NIBP Front<br />
Detail View<br />
Adjust<br />
VR 81<br />
Here