Calibration Procedure Diagnostics<br />

6. Adjust R81 potentiometer and verify the linearity accordingly. See figure below.<br />

Static Calibration Setup<br />

Digital Mercury Manometer<br />

Bulb<br />

150 mmHg<br />

FIGURE 6-13<br />

Bulb<br />

Valve<br />

Manually<br />

Squeeze<br />

Bulb<br />

T - Fitting<br />

NIBP Module<br />

693cc<br />

Test Chamber<br />

P/N 0138-00-0001-01<br />

Specification: = 150mmHg +/- 0.1%<br />

Previous Menu<br />

Overpressure Voltage Setpoint Static Pressure Calibration<br />

Static Pressure Calibration<br />

Motor Pump Test Transducer: 150.0 mmHg<br />

Leak Test<br />

Dump Test<br />

Bleed Rate Test<br />

Overpressure Test<br />

Abort Test<br />

7. Reassemble unit and verify the following static pressure points to ensure proper<br />

calibration.<br />

Specification: 30 mmHg, 100 mmHg, 150 mmHg, 190 mmHg and 250<br />

mmHg. +/-3 mmHg.<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Service Manual 0070-10-0441 6 - 11<br />

VR 81<br />

Select to return to previous menu.<br />

NIBP Test<br />

<strong>Passport</strong> 2<br />

NIBP Front<br />

Detail View<br />

Adjust<br />

VR 81<br />

Here