Datascope Passport - Mindray
Datascope Passport - Mindray
Datascope Passport - Mindray
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
SpO 2 Interface Board (Nellcor, Nell-3 and Masimo) Theory of Operation<br />
2.16 SpO 2 Interface Board (Nellcor, Nell-3 and Masimo)<br />
The SpO2 Interface board provides the communications path for the two-way data flow<br />
between the SpO2 modules and the <strong>Passport</strong> 2 CPU Control Module. The SpO2 module<br />
analyzes the analog input from the SpO2 sensor and outputs the data in digital format via a<br />
serial digital interface.<br />
The SpO2 Interface board addresses the different power requirements, as well as the<br />
different connector pin assignments for each of the three SpO2 modules.<br />
The SpO2 interface board provides SCI UART host communication to the SpO2 modules. It<br />
serves simply as an adapter. The signals pass through unchanged.<br />
The SpO2 interface board will be specifically configured for the SpO2 module it is<br />
interfacing with. This may include different jumper settings, different feedback resistor<br />
options, custom ribbon cables, etc.<br />
Patient isolation is provided on the <strong>Passport</strong> 2 CPU Control Module, so it is not needed on<br />
this SpO2 interface board.<br />
2 - 54 0070-10-0441 <strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Service Manual