Preventive Maintenance Care And Cleaning Of The Monitor<br />

7.7 Care And Cleaning Of The Monitor<br />

The monitor enclosure may be cleaned with a mild soap and water solution, isopropyl<br />

alcohol, 10% bleach and water solution or ammoniated window cleaner. Apply cleaning<br />

solution to the cloth, not directly onto the monitor. DO NOT apply large amounts of liquid.<br />

DO NOT use abrasive cleaning agents or organic solvents.<br />

WARNING: Do not clean the monitor while it is on and/or plugged in.<br />

To prevent scratches on the front panel display screen, blow or carefully brush dust and dirt<br />

particles with a soft sponge moistened with cleaner solution or a fine, soft-hair brush. DO<br />

NOT use abrasive cleaning materials. Fingerprints and stains may be removed by using a<br />

liquid lens cleaner and a soft cloth. DO NOT wipe a dry screen or use alcohol or chlorinated<br />

hydrocarbon solvents.<br />

7.8 Care and Cleaning of SpO 2 Sensor<br />

NOTE: Refer to the individual instruction sheets packaged with<br />

each sensor.<br />

1. Check sensors and cables for wear or damage daily. Replace as required.<br />

2. Check for proper operation of the spring mechanism on reusable adult sensors.<br />

3. Sensors should be cleaned before and after use on each patient.<br />

4. Clean and disinfect the sensors by wiping the patient contact area with a soft cloth<br />

dampened with a mild soap and water solution or isopropyl alcohol. Hydrogen<br />

peroxide can be used to remove dried blood on all accessible surfaces.<br />

5. DO NOT immerse sensors. Let sensors dry completely before using.<br />

CAUTION: When cleaning sensors do not use excessive amounts of<br />

liquid. Wipe the sensor with a soft cloth, dampened with the<br />

cleaning solution.<br />

7.9 Cleaning CO 2 Sensors, Adapters And Sampling<br />

Components<br />

Oridion CO2 patient monitoring accessories are designed for single patient use and should<br />

not be cleaned or reused.<br />

<strong>Passport</strong> 2®/<strong>Passport</strong> 2 LT Service Manual 0070-10-0441 7 - 3