Operation4.6Electronics, Ground Connection, EarthingThe sensor housings are connected to signal ground and to supply ground by capacitors.Signal ground and supply ground are dc-insulated for the 24 V option.Mounting the sensor in a metal holder establishes a sufficient ground connection.An additional groundconnection is required for small, remotely located targets. Connect the target to thesensor housing for this.5.Operation5.1i5.2Starting UpConnect the display/output devices through the signal output socket, see Chap. 4.5, before connectingthe device to the power supply and switching on the power supply.Allow the measuring system to warm up for about one minute before the first measurement or calibration.Basic SettingsThe gain and zero point of the measuring channel are adjusted with the ’gain’ and ’zero’ trimmer potentiometers,see Fig. 6, (setting range approx. 20 turns per potentiometer).The end settings at the left and right stops are recogniceable by a slight click.English<strong>capaNCDT</strong> <strong>6100</strong>Page 53