OperationStep 3Repeat steps 1 and 2 about three times until a stable potentiometer setting results.Replace the housing cover.If the sensor face and the target are mismatched (not parallel), there will be a loss in the range and a shift inthe zero point depending on the tilt.Curved target surfaces cause linearity reductions if the distance between the sensor and the target is small.Linearity and sensitivity deviations also occur with a too small target dimensions.10 V2Outputvoltage0 V10 % Measuring range 100 %SensorFig. 7 Calibration stepsTarget1 = Start of measuring range2 = End of measuring rangeController with option I (current output):The values-- 4 mA (start of measuring range) and-- 20 mA (end of measuring range) are set for systems with current output.<strong>capaNCDT</strong> <strong>6100</strong>Page 56