12.07.2015 Aufrufe

capaNCDT 6100 - Micro-Epsilon

capaNCDT 6100 - Micro-Epsilon

capaNCDT 6100 - Micro-Epsilon

MEHR ANZEIGEN
WENIGER ANZEIGEN
  • Keine Tags gefunden...

Sie wollen auch ein ePaper? Erhöhen Sie die Reichweite Ihrer Titel.

YUMPU macht aus Druck-PDFs automatisch weboptimierte ePaper, die Google liebt.

Operation5.5Synchronization in Multichannel SystemsSeveral measuring systems in the <strong>capaNCDT</strong> <strong>6100</strong> series can be operated simultaneously as multichannelsystems.Synchronization of the measuring systems avoids interference between the sensors:Plug SC30 synchronization cable (accessory) in the SYN OUT (synchronization output) socket in controller1.Plug the other end on the cable into the SYN IN (synchronization input) in controller 2.-- The oscillator of controller 2 switches automatically to synchronization mode, i.e. it acts as slave to themaster oscillator in controller 1.-- The influence of unsufficient grounded targets will be excepted.Synchronise several measuring systems with SC30 cable if necessary.Controller 1 Controller 2 Controller 3SYNOUTSYNINSC30SYNOUTSYNINSC30Fig. 9 Synchronization in multichannel systemsiAutomatic synchronization, each controller can work as master.Bending radius SC30:-- 30 mm (one time)-- 60 mm (alternating)<strong>capaNCDT</strong> <strong>6100</strong>Page 58

Hurra! Ihre Datei wurde hochgeladen und ist bereit für die Veröffentlichung.

Erfolgreich gespeichert!

Leider ist etwas schief gelaufen!