3.1. CHEMISTRY OF THE CLARK SENSOR 15<br />

face. When the negative bias voltage is further increased, the current output<br />

<strong>of</strong> the electrode increases rapidly due to other reactions. If a fixed voltage in<br />

the plateau region (<strong>for</strong> instance, -0.7V) is applied to the cathode, the current<br />

output <strong>of</strong> the electrode can be linearly calibrated to the dissolved oxygen<br />

(short termed DO). It should be noted that the current is proportional not<br />

to the actual concentration but to the activity or equivalent partial pressure<br />

<strong>of</strong> dissolved oxygen, which is <strong>of</strong>ten referred to as oxygen tension. A fixed<br />

voltage between -0.6 and -0.8 V (due to the cell potential, and the onset <strong>of</strong><br />

other processes) is usually selected as the polarization voltage when using<br />

Ag/AgCl as the reference electrode.<br />

Figure 3.2: Polarogram: Current to voltage at different oxygen tensions and<br />

calibration obtained at a fixed polarization voltage. Reproduced<br />

from [24]. The shape <strong>of</strong> the curves indicates the nature <strong>of</strong> the<br />

substance in the sample that is being reduced; where the height<br />

is equivalent to the concentrations <strong>of</strong> oxygen. Also note that<br />

convention usually have -V along the x-axis and not V, as is<br />

otherwise the norm.<br />

The cathode, the reference electrode, and the electrolyte are separated<br />

from the measurement medium by a membrane, which is permeable to the<br />

dissolved gas but not to most <strong>of</strong> the ions and other species. When at the<br />

same time most <strong>of</strong> the mass transfer resistance is confined in the membrane,<br />

the electrode system can measure oxygen tension in various liquids. This is<br />

the basic operating principle <strong>of</strong> the membrane covered DO probe[25].<br />

Electrode Reactions: The reaction can proceed as follows (a 4e − reaction<br />

with a Ag/AgCl electrode is used as an example)<br />

Cathodic reaction: