Operating Instructions - VEGAPULS 54K enamel
Operating Instructions - VEGAPULS 54K enamel
Operating Instructions - VEGAPULS 54K enamel
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
24 101-EN-041227<br />
Technical data<br />
6 Technical data<br />
6.1 Technical data<br />
Power supply<br />
Power supply<br />
- four-wire sensor 24 V DC (20 … 72 V DC)<br />
230 V AC (20 … 250 V AC), 50/60 Hz<br />
fuse 0.2 A TR<br />
- two-wire sensor 24 V DC (14 … 36 V DC)<br />
- two-wire Ex ia sensor 24 V DC (14 … 29 V DC)<br />
- two-wire Ex d ia sensor 24 V DC (20 … 36 V DC)<br />
Current consumption<br />
- four-wire sensor max. 60 mA<br />
- two-wire sensor max. 22.5 mA<br />
Power consumption<br />
- four-wire sensor max. 200 mW, 1.2 VA<br />
- two-wire sensor 55 … 810 mW<br />
Load<br />
- four-wire sensors max. 500 Ohm<br />
- two-wire sensors see load diagram<br />
Ω<br />
1000<br />
975<br />
900<br />
800<br />
720<br />
700<br />
670<br />
600<br />
500<br />
400<br />
max. load Ex ia<br />
max. load non-Ex<br />
max. load Ex d ia<br />
Non-Ex and Ex ia<br />
max. voltage limit<br />
non-Ex and<br />
Ex d ia sensors<br />
max. Spannungsgrenze<br />
Ex ia-Sensoren<br />
300<br />
250<br />
200<br />
Load resistance<br />
(HART<br />
100<br />
0<br />
15 19,5<br />
25,5 29 36<br />
14 20 25 30 35<br />
® min. voltage limit when using the HART<br />
and<br />
VEGACONNECT)<br />
®<br />
adjustment resistance:<br />
- non-Ex and Ex ia sensors<br />
- Ex d ia sensors<br />
Exd ia<br />
<strong>VEGAPULS</strong> <strong>54K</strong> <strong>enamel</strong> 39<br />
V