Operating Instructions - VEGAPULS 54K enamel
Operating Instructions - VEGAPULS 54K enamel
Operating Instructions - VEGAPULS 54K enamel
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Ex-technical data<br />
Comprehensive data in the safety instructions manual (yellow binder)<br />
Connection cables<br />
Technical data<br />
Two-wire sensors power supply and signal via one<br />
two-wire cable<br />
Four-wire sensors power supply and signal separated<br />
Electrical connection - spring-loaded terminal connection<br />
(max. 2.5 mm 2 )<br />
- plug connection (secure by cable entry)<br />
Cable entry<br />
-ia terminal compartment 1 … 2 x M20 x 1.5 (cable-ø 5 … 9 mm)<br />
or 2 x 1 / 2 “ NPT (cable-ø<br />
3.6 … 8.6 mm or 0.12 … 0.34 inch)<br />
- Ex d terminal compartment 1 x 1 / 2 “ NPT EEx d (cable-ø<br />
3.6 … 8.6 mm or 0.12 … 0.34 inch)<br />
Ground connection max. 4 mm 2<br />
Materials<br />
Housing PBT (Valox) or<br />
Aluminium die casting (GD-AlSi 10 Mg)<br />
EEx d connection compartment Aluminium mould casting (GK-AlSi 7 Mg)<br />
Process fitting/Antenna <strong>enamel</strong>/PTFE<br />
Antenna seal special PTFE seal for <strong>enamel</strong>led vessels<br />
Weights<br />
DN 150 27 kg<br />
DN 200 41 kg<br />
ANSI 6“ 33 kg<br />
ANSI 8“ 48 kg<br />
CE conformity<br />
<strong>VEGAPULS</strong> series 50 radar sensors meet the protective regulations of EMC (89/336/EWG)<br />
and NSR (73/23/EWG) and R & TTE directive (1999/5/EC).<br />
Conformity was judged acc. to the following standards:<br />
EN 300 683-1: 1997<br />
EN 300 440-1: 1995<br />
I-ETS 300-440<br />
Expert opinion No. 0043052-01/SEE, Notified Body No. 0499<br />
EN 61 326: 1997/A1: 1998 (EMC Emission/Susceptibility)<br />
EN 61 010 - 1: 1993 (NSR)<br />
EN 50 020: 1994 (ATEX)<br />
EN 50 018: 1994<br />
EN 50 014: 1997<br />
42 <strong>VEGAPULS</strong> <strong>54K</strong> <strong>enamel</strong><br />
24 101-EN-041227