Operating Instructions - VEGAPULS 54K enamel
Operating Instructions - VEGAPULS 54K enamel
Operating Instructions - VEGAPULS 54K enamel
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
24 101-EN-041227<br />
Technical data<br />
Measurement characteristics 1)<br />
(typical values under reference conditions, all statements relate to the nominal measuring<br />
range)<br />
Min. span between<br />
full and empty > 10 mm (recommended > 50 mm)<br />
Frequency 5.8 GHz (USA 6.3 GHz)<br />
Intervals<br />
- two-wire sensor (4 … 20 mA) 1 s<br />
- four-wire sensor 0.5 s<br />
Beam angle (at –3 dB)<br />
- <strong>VEGAPULS</strong> 54 with DN 150 20°<br />
- <strong>VEGAPULS</strong> 54 with DN 200 16°<br />
Influence of the process temperature cannot be measured at 0 bar; at 5 bar 0.004 %/<br />
10 K;<br />
at 40 bar 0.03 %/10 K<br />
Influence of the process pressure 0.0265 %/bar<br />
Ambient conditions<br />
Ambient temperature on the housing -20°C … +60°C<br />
Process temperature (flange temp.) -40°C … +200°C<br />
bar<br />
16<br />
-40 0 175 200<br />
Process pressure/Vessel pressure see diagram<br />
Storage and transport temperature -40°C … +80°C<br />
Protection IP 66 and IP 67<br />
Protection class<br />
- two-wire sensor II<br />
- four-wire sensor I<br />
Overvoltage category III<br />
1) Similar to DIN 16 086, reference conditions acc. to IEC 770, e.g.<br />
temperature 15 °C … 35 °C; moisture 45 % … 75 %; pressure 860 mbar … 1060 mbar<br />
<strong>VEGAPULS</strong> <strong>54K</strong> <strong>enamel</strong> 41<br />
˚C