31.12.2013 Views

DRAFT Recommended Practice for Measurements and ...

DRAFT Recommended Practice for Measurements and ...

DRAFT Recommended Practice for Measurements and ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

1/29/98 77 C95.3-1991 Revision — 2 nd Draft<br />

10/98 Draft<br />

frequencies. There<strong>for</strong>e, fiber optically coupled instruments are preferable at frequencies<br />

below 300 MHz.<br />

RFI also becomes a significant problem at lower frequencies, since it is more difficult to<br />

shield readout electronics <strong>and</strong> cables at frequencies below about 500 MHz. An<br />

operational test of the system can be made by totally shielding the sensor tip of an<br />

electric-field probe with metal foil. This procedure allows the determination of the<br />

existence of RFI or capacitive coupling between the cable <strong>and</strong> readout <strong>and</strong> nearby<br />

radiating objects during the measurement. Similarly, reorienting the probe leads in the<br />

field, with the probe sensing element fixed, should not alter the reading of the meter<br />

appreciably (±6 dB) if RFI is not problematic. It is not uncommon to find large errors in<br />

readings from older designs of isotropic instruments when they are used to measure<br />

electric <strong>and</strong> magnetic fields below about 1000 MHz, especially when the readout device<br />

<strong>and</strong> probe cable are exposed to a field of the same magnitude as that which illuminates<br />

the probe [B16, B109]. This often is due to faulty RFI gaskets in the readout device or<br />

poor system RFI design. Fiber-optic cables, high-resistance cables, or double-shielded<br />

coaxial <strong>and</strong> signal cables can be used by designers of hazard survey instruments to<br />

connect the probe to the readout device, in order to minimize the effects of RFI. Also, a<br />

self-contained (active antenna) probe with integrated antenna, sensor, <strong>and</strong> readout in the<br />

same housing can minimize the problem of RFI otherwise induced in cables or readout<br />

[B19].<br />

Instrument linearity as a function of the peak-to-average power ratio of amplitudemodulated<br />

fields should be considered when per<strong>for</strong>ming measurements in an unknown<br />

situation. Pulse duty factor or amplitude-modulation wave<strong>for</strong>m can be quantified through<br />

the use of a conventional antenna with a coaxial output, e.g., log periodic, dipole, <strong>and</strong> a<br />

coaxial attenuator <strong>and</strong> diode detector. The detector output can be fed via coaxial cable<br />

into an oscilloscope of appropriate b<strong>and</strong>width. The duty factor of pulsed sources <strong>and</strong> the<br />

wave<strong>for</strong>m of the amplitude modulation can then be observed, <strong>and</strong> the ratio of the peak to<br />

average power or energy density can be ascertained. It is not necessary to maintain<br />

square-law operation of the detector <strong>for</strong> the evaluation of 100% rectangular amplitudemodulated<br />

wave<strong>for</strong>ms. For other pulse modulated wave<strong>for</strong>ms, one can only be assured<br />

that both the peak <strong>and</strong> average field strengths are correct if the detector is operated in the<br />

square law region [B118]. Also, with square law detection, the ratio of the peak to<br />

average equivalent power density, (E 2 or H 2 ) can be quantified.<br />

Measuring the effects of a slowly time-varying field generated by a rotating radar antenna,<br />

or a microwave oven with a mode stirrer, can be accommodated by use of the<br />

appropriate time constant of the readout instrument. Some instruments provide ‘’fast’’<br />

<strong>and</strong> ‘’slow’’ time constants. Any observable periodic variation of the field strength, as<br />

indicated by the instrument in the fast mode of operation, should be considered as lowfrequency<br />

field modulation <strong>and</strong> the slow time constant should be used. Care should be<br />

taken to avoid mistaking slow variations in reflected signals (due to objects moving in the<br />

field) <strong>for</strong> a true time-varying source modulation. This implies that any movement of the<br />

operator, probe, or surrounding objects, should be eliminated during the measurement<br />

procedure where time-varying fields are involved.<br />

An ellipticity (non-isotropic reception condition) in the receiving pattern of multi-axis or<br />

isotropic probes is always present to a certain degree, (typically 2% to 20%). Only if the<br />

ellipticity of an instrument is accounted <strong>for</strong>, can highly accurate data be taken. In fields<br />

where components do not exist along the direction of the probe h<strong>and</strong>le, a rotation of the<br />

probe about the h<strong>and</strong>le axis should be per<strong>for</strong>med during a survey (if the surveyor desires<br />

Copyright © 1998 IEEE. All rights reserved. This is an unapproved IEEE St<strong>and</strong>ards Draft,<br />

subject to change.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!