31.12.2013 Views

DRAFT Recommended Practice for Measurements and ...

DRAFT Recommended Practice for Measurements and ...

DRAFT Recommended Practice for Measurements and ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

1/29/98 92 C95.3-1991 Revision — 2 nd Draft<br />

10/98 Draft<br />

probe has a response time of the order of a few milliseconds, continuous line scans of<br />

the internal E-fields may be dynamically recorded by means of a robot that generates<br />

position data as the probe is moved along a path. Extensive data in an object can, thus,<br />

be plotted in a relatively short period of time, <strong>and</strong> the possibility of missing a local peak is<br />

reduced [Bassen, et al., 1977].<br />

There are several inherent sources of error associated with the use of implantable E-field<br />

probes <strong>for</strong> SAR measurements. Regardless of the quality of the specific probe used,<br />

calibration (in terms of absolute field strength in high water-content biological tissue or<br />

tissue simulant) is difficult (see below). Large gradients in the internal E-fields <strong>and</strong><br />

imprecise knowledge of the conductivity <strong>and</strong> mass density of the biological tissue or<br />

tissue simulant add additional degrees of uncertainty. Although commercially available<br />

implantable probes are available, custom-designed probes are frequently developed,<br />

evaluated <strong>and</strong> calibrated by the developer <strong>and</strong>/or user. There<strong>for</strong>e, instrument<br />

per<strong>for</strong>mance limitations, as well as errors introduced during the measurement<br />

procedure, should be understood by the user <strong>and</strong> steps taken to minimize errors<br />

introduced by these factors. Detailed discussions on the evaluation, calibration <strong>and</strong> use<br />

of implantable probes have been published [Bassen, et al., 1977, Stuchly, 1987]. In<br />

summary, SAR measurement uncertainties of at least ±2 dB are to be expected, even<br />

under optimum measurement conditions.<br />

The high sensitivity of the E-field probes make them ideally suited <strong>for</strong> the measurement<br />

of SARs associated with low-power (of the order of one watt or less) localized sources<br />

such as h<strong>and</strong>-held radio transceivers, e.g., cellular <strong>and</strong> personal communications<br />

equipment. The low power output <strong>and</strong> small size of the antenna make thermal<br />

measurements extremely difficult [Balzano, et al., 1978a; Balzano, et al., 1979; Balzano,<br />

et al., 1978]. Increasing the power of such sources by a factor of ten times so that<br />

thermal techniques could be used would entail a substantial modification of the device to<br />

the point of being non-representative of an actual transceiver. Since the exposure of<br />

concern from such low power sources localized sources is within about 5 cm of the<br />

antenna, accurate positioning of the sensing probe is crucial <strong>for</strong> per<strong>for</strong>ming repeatable<br />

measurements. The probe positioning should be per<strong>for</strong>med by machine, e.g., a robot,<br />

rather than by h<strong>and</strong>.Simulated human tissue used <strong>for</strong> such measurements may vary<br />

from case to case depending on the specific device being evaluated. For cellular<br />

phones, <strong>for</strong> example, the head <strong>and</strong> upper part of the torso would usually be sufficient<br />

[Balzano, 1995], while a<br />

150 MHz two-way radio strapped to the belt with the antenna operated by remote<br />

switching would require a full-size human phantom [Fujimoto, 1994].<br />

The conductivity of the tissue simulant must be correct <strong>for</strong> the frequency being tested.<br />

The mixing of such materials <strong>and</strong> the measurement of their corresponding electrical<br />

properties present substantial challenges to achieve accuracy <strong>and</strong> repeatability. It is not<br />

possible to use a single <strong>for</strong>mulation over a wide frequency range, e.g., more than an<br />

octave, without running into relatively large departures (±5 %) from published<br />

conductivities <strong>for</strong> biological tissues. To obtain repeatable results (±3 %) it is advisable to<br />

restrict the frequency b<strong>and</strong> <strong>and</strong> to purchase the primary materials from the same<br />

supplier. Well documented mixing procedures should also be adhered to, e.g., accurate<br />

weights of the components, temperature of the liquids during mixing, length of the mixing<br />

time, RPM of the stirring device. Measurement of the dielectric properties of tissue<br />

simulants is also difficult to per<strong>for</strong>m accurately. Acceptable results can be obtained<br />

Copyright © 1998 IEEE. All rights reserved. This is an unapproved IEEE St<strong>and</strong>ards Draft,<br />

subject to change.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!