26.01.2014 Views

NewTom VG User Manual rev 4.0 - Image Works

NewTom VG User Manual rev 4.0 - Image Works

NewTom VG User Manual rev 4.0 - Image Works

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>VG</strong>i <strong>User</strong> <strong>Manual</strong><br />

Rev. 1.0 - February, 15 2009


NOTES:<br />

This document is provided for the own use of the operator of the equipment.<br />

QR s.r.l. reserves the right to change the contents of this manual without notice.<br />

This document may not, in whole or in part, be modified, copied, reproduced, distributed, translated, stored<br />

on magnetic or optical media and published, over networks, electronic bulletin boards, web sites or other online<br />

services, without the express written permission of QR s.r.l.<br />

The original version of this manual has been written in English language.<br />

NEWTOM <strong>VG</strong>i is a commercial trademark of QR s.r.l.<br />

All other products and brand names are registered trademarks or trademarks of their respective companies.<br />

NEWTOM <strong>VG</strong>i is manufactured and distributed by:<br />

QR srl<br />

Via Silvestrini, 20<br />

37135 Verona<br />

Italy<br />

Phone: +39 045 8202727<br />

Fax +39 045 8203040<br />

e-mail: info@qrverona.it<br />

www.qrverona.it<br />

All rights reserved.<br />

Rev. 1.0 - 02/15/2009 Page 1 of 72


MANUFACTURER’S NOTICE ABOUT MEDICAL DEVICES:<br />

The medical device described in this manual, composed by Scan Unit (gantry), patient support (patient table)<br />

and Control, display and computation unit (Main <strong>Works</strong>tation) as delivered and configured by the<br />

manufacturing and servicing personnel, is a medical radiological device subject to the safety requirement<br />

stated by both international standards and local regulation.<br />

Any change, modification, update on either hardware 1 or software 2 , on the device and operating settings as<br />

released form factory and as installed (and in any case different from what has been specified on the<br />

device's attached documentation), can cause the partial or total device impossibility to operate as expected.<br />

Moreover its safety features can be changed with increased hazard for the patient, operator and surrounding<br />

environment.<br />

So, in case that the final user need to modify or change any of the above specified items, he/she has to be<br />

allowed in written by QR s.r.l.<br />

Any act outside of what has been stipulated on this notice will cancel the warranty. Moreover the person<br />

and/or organization will be liable for any damage and/or incident and/or deterioration in the state of health of<br />

a patient, user or other person and/or the surrounding environment.<br />

1 Adding a memory expansion, a new board on the connection bus, a new printer, the replacement of the graphic adapter is a<br />

significant change.<br />

2 Including the Operating System an applications already installed when the device has been delivered and/or installed. Operating<br />

system automatic updates, changes on the network connection parameters, modification and/or addition and/or removal of hardware<br />

interface software (device driver) and/or system services (example: file sharing) and/or applications, constitute a significant change.<br />

Rev. 1.0 - 02/15/2009 Page 2 of 72


Index<br />

INDEX<br />

1 INTRODUCTION TO THIS MANUAL........................................................................................................ 5<br />

1.1 CONTENTS................................................................................................................................................. 5<br />

1.2 STRUCTURE............................................................................................................................................... 5<br />

1.3 STYLE CONVENTIONS ................................................................................................................................. 6<br />

1.4 ABOUT USING THIS MANUAL ........................................................................................................................ 6<br />

2 ABOUT SAFETY ....................................................................................................................................... 7<br />

2.1 GENERAL SAFETY RULES ............................................................................................................................ 7<br />

2.2 TURNING THE DEVICE OFF........................................................................................................................... 8<br />

2.3 EMERGENCY SHUTDOWN ............................................................................................................................ 9<br />

2.4 PATIENT'S AND USER'S SAFETY GUIDELINES ............................................................................................... 10<br />

2.4.1 Patient positioning ........................................................................................................................... 10<br />

2.4.2 Patient's exit .................................................................................................................................... 11<br />

2.4.3 Protection from ionizing radiation.................................................................................................... 11<br />

2.5 LASER EXPOSURE .................................................................................................................................... 12<br />

2.6 SAFETY NOTES......................................................................................................................................... 12<br />

3 SAFETY AND MAINTENANCE OF THE DEVICE.................................................................................. 14<br />

3.1 INSTALLATION REQUIREMENTS .................................................................................................................. 14<br />

3.2 GUIDELINES FOR SAFETY .......................................................................................................................... 15<br />

3.3 CHANGES IN THE DEVICE .......................................................................................................................... 15<br />

3.4 DEVICE MAINTENANCE .............................................................................................................................. 16<br />

3.5 CLEANING AND DISINFECTION.................................................................................................................... 17<br />

3.5.1 Chemical agents to avoid................................................................................................................ 17<br />

3.5.2 Cleaning .......................................................................................................................................... 18<br />

3.5.3 Disinfection...................................................................................................................................... 18<br />

3.5.4 Sterilization...................................................................................................................................... 18<br />

3.6 TRANSPORT AND STORAGE....................................................................................................................... 18<br />

3.7 DEVICE DISPOSAL..................................................................................................................................... 19<br />

3.7.1 Information for the device owner..................................................................................................... 19<br />

3.7.2 Information for the Disposing/Recycle centers ............................................................................... 19<br />

4 GETTING STARTED…............................................................................................................................ 20<br />

4.1 AN INTRODUCTION TO THE SYSTEM............................................................................................................ 20<br />

4.2 WORKING PRINCIPLE ................................................................................................................................ 20<br />

4.3 OVERALL VIEW ......................................................................................................................................... 21<br />

4.4 SCANNER UNIT ......................................................................................................................................... 22<br />

4.4.1 Controls placed on the scanner unit................................................................................................ 22<br />

4.4.2 Controls on the head support keyboard .......................................................................................... 23<br />

4.5 CONTROL BOX.......................................................................................................................................... 24<br />

4.5.1 Main switch...................................................................................................................................... 24<br />

4.5.2 Input panel....................................................................................................................................... 24<br />

4.6 STANDARD ACCESSORIES ......................................................................................................................... 25<br />

4.6.1 Cables ............................................................................................................................................. 26<br />

4.7 OPTIONAL ACCESSORIES .......................................................................................................................... 26<br />

4.8 SYSTEM STARTUP .................................................................................................................................... 27<br />

4.9 SYSTEM SHUTDOWN................................................................................................................................. 27<br />

5 PRELIMINARY PROCEDURES .............................................................................................................. 28<br />

5.1 X-RAY SOURCE CONDITIONING.................................................................................................................. 28<br />

5.2 DAILY CHECK........................................................................................................................................... 28<br />

5.3 BLANK ACQUISITION.................................................................................................................................. 29<br />

5.3.1 Invalidate blank acquisition ............................................................................................................. 29<br />

5.4 BEAM LIMITER TEST .................................................................................................................................. 30<br />

5.5 COMPLETELY LIFT DOWN THE ROTATING ARM............................................................................................. 30<br />

6 SCANNING .............................................................................................................................................. 31<br />

6.1 SCANNING A PATIENT................................................................................................................................ 31<br />

Rev. 1.0 - 02/15/2009 Page 3 of 72


Index<br />

6.1.1 Preparing the patient ....................................................................................................................... 31<br />

6.1.2 Positioning the patient and starting a new scan.............................................................................. 33<br />

6.2 SCANNING A DENTURE.............................................................................................................................. 35<br />

6.2.1 Preliminary operations..................................................................................................................... 35<br />

6.2.2 Denture positioning ......................................................................................................................... 35<br />

7 QUALITY ASSURANCE.......................................................................................................................... 36<br />

7.1 PHANTOM POSITIONING ............................................................................................................................ 36<br />

7.2 SAMPLE IMAGES....................................................................................................................................... 38<br />

7.3 STORAGE OF QA DATA............................................................................................................................. 38<br />

8 TROUBLESHOOTING............................................................................................................................. 39<br />

8.1 COMMUNICATIONS PROBLEM..................................................................................................................... 39<br />

8.2 ROTATING ARM......................................................................................................................................... 39<br />

8.3 IMAGE ACQUISITION .................................................................................................................................. 39<br />

9 IEC61223: ACCEPTANCE TEST............................................................................................................ 40<br />

9.1 POSITIONING OF THE PATIENT SUPPORT [PARAGRAPH 5.1].......................................................................... 40<br />

9.2 PATIENT POSITIONING ACCURACY [PARAGRAPH 5.2]................................................................................... 40<br />

9.2.1 Axial patient positioning accuracy [paragraph 5.2.1] ...................................................................... 40<br />

9.2.2 Sagittal and coronal patient positioning light [paragraph 5.2.2] ...................................................... 40<br />

9.3 TOMOGRAPHIC SECTION THICKNESS [PARAGRAPH 5.3]............................................................................... 40<br />

9.3.1 Tomographic section thickness for axial scanning [paragraph 5.3.1] ............................................. 40<br />

9.3.2 Tomographic section thickness for helical scanning [paragraph 5.3.2] .......................................... 40<br />

9.4 DOSE [PARAGRAPH 5.4]............................................................................................................................ 40<br />

9.5 NOISE, MEAN CT NUMBER AND UNIFORMITY [PARAGRAPH 5.5] ................................................................... 40<br />

9.6 SPATIAL RESOLUTION [PARAGRAPH 5.6] .................................................................................................... 40<br />

9.7 IEC 61223-3-5 REFERENCE TABLE........................................................................................................... 41<br />

APPENDIX A: TECHNICAL REFERENCE .................................................................................................. 43<br />

APPENDIX B: STANDARDS CONFORMITY .............................................................................................. 63<br />

APPENDIX C: LABELS ................................................................................................................................ 65<br />

Rev. 1.0 - 02/15/2009 Page 4 of 72


Introduction to this manual<br />

1 Introduction to this manual<br />

1.1 Contents<br />

This manual was created as a mean to provide information and instructions about using a <strong>NewTom</strong> <strong>VG</strong>i<br />

device.<br />

Dedicated software routines (scan, data processing, reporting, document management) and their user<br />

instructions are described in the attached Software <strong>Manual</strong>.<br />

"<strong>User</strong> <strong>Manual</strong>" and "Software <strong>Manual</strong>" should be read and clearly understood in every part before starting<br />

using the device.<br />

It is suggested to store this manual along with any other documentation and to use it whenever a new staff<br />

member has to be trained for using the device.<br />

1.2 Structure<br />

This manual is divided in the following chapters:<br />

• Chapter 1 (Introduction to this manual):<br />

it provides information about contents, structure and conventions inside the document.<br />

• Chapter 2 (About safety):<br />

information about user's and patients' safety and primary procedures regarding the use of the<br />

device.<br />

• Chapter 3 (Safety and maintenance of the device):<br />

information about safety requirements and the device maintenance.<br />

• Chapter 4 (Getting started…):<br />

gives an overall description of the system and of its main parts.<br />

• Chapter 5 (Preliminary procedures):<br />

illustrates the procedure for a proper initialization of the device.<br />

• Chapter 6 (Scanning):<br />

describes the process for patient's positioning and scanning.<br />

• Chapter 7 (Quality assurance):<br />

gives instructions about executing the Quality Assurance process.<br />

• Chapter 8 (Troubleshooting):<br />

a list of faults and suggestions about their resolution.<br />

• Chapter 9 (IEC61223: Acceptance Test):<br />

procedures for testing as required by IEC61223 standard.<br />

Besides these chapters, you can find the appendices:<br />

• APPENDIX A: Technical Reference<br />

• APPENDIX B: Standards conformity<br />

• APPENDIX C: Labels<br />

Rev. 1.0 - 02/15/2009 Page 5 of 72


Introduction to this manual<br />

1.3 Style conventions<br />

Primary information about safety and possible notes are included in this manual in the following form:<br />

DANGER:<br />

It alerts for a potential danger which may cause injury or death of a person.<br />

WARNING:<br />

It signals the presence of a potential danger which may damage the device.<br />

NOTE:<br />

It gives information not regarding the device and/or the user's and/or a<br />

patient's safety.<br />

1.4 About using this manual<br />

This manual is written in such a way to provide the user all the know-how that is necessary for the device,<br />

the user's and a patient's safety. To this purpose, chapters regarding people and device safety must be read<br />

carefully, getting confident with the procedures described there.<br />

Rev. 1.0 - 02/15/2009 Page 6 of 72


About safety<br />

2 About safety<br />

This chapter includes information about safety that must be learnt by the user before using the device.<br />

For a proper use of a <strong>NewTom</strong> <strong>VG</strong>i device, the operator needs to have in-depth knowledge of the mode of<br />

operation and of the related controls.<br />

The table below describes the symbols found on the device labels.<br />

Symbol IEC Description<br />

~ 417-5032 Alternate current<br />

348 Caution, consult the attached documentation<br />

417-5019 Protection ground<br />

417-5017 Functional ground<br />

N 445 Connection point of the neutral wire<br />

L<br />

Connection point of the Line<br />

417-…<br />

878-02-02<br />

Type B (Applied part)<br />

ISO 361<br />

Ionizing Radiation<br />

Directive<br />

2002/96/CE<br />

WEEE - Waste from Electrical and Electronic<br />

Equipment Symbol<br />

2.1 General safety rules<br />

In order to guarantee people's safety, please always follow the instructions inside this manual, especially the<br />

ones about functional tests, mechanical and electrical safety and protection against ionizing radiation<br />

exposure. Please refer to:<br />

Chapter 2 "About safety"<br />

Chapter 3 "Safety and maintenance of the device"<br />

Chapter 6 "Scanning"<br />

Rev. 1.0 - 02/15/2009 Page 7 of 72


About safety<br />

Local laws<br />

Carefully follow every requirements about installation and maintenance of the device. Please refer to the<br />

local laws whenever these may be more strict than the instructions inside this manual.<br />

Switching on and off of the device<br />

Turning the device on and off must be performed according to the procedures described inside this manual<br />

(see 2.2 "Turning the device off" and 4.8 "System startup").<br />

Emergency stop<br />

In case of an emergency, follow the instructions referring to the device emergency stop, see 2.3 "Emergency<br />

shutdown".<br />

Maintenance time lag<br />

Please be sure that the maintenance controls that are described in 3.4 "Device maintenance" are performed.<br />

During patient positioning<br />

Before activating the positioning laser always warn the patient to close his/her eyes.<br />

While scanning a patient<br />

NEVER leave the system without a supervisor during the execution of an exam. Always watch the patient all<br />

along the scan process.<br />

Artifacts<br />

Please contemplate the repeating of a scan ONLY if suspect or significant artifact are evident in a patient's<br />

images, or if a patient's position appreciably changed during the exam.<br />

WARNING:<br />

Every operator must have a deep knowledge of the system operative and<br />

environmental features, he/she also must know the procedures to follow in<br />

case of emergency.<br />

WARNING:<br />

Never use the device without the supervision of an operator.<br />

NOTE:<br />

Please consider the realization of an audio/video communication system between the<br />

operator and the patient, especially when the operator controls the device from a<br />

protected and remote location.<br />

2.2 Turning the device off<br />

Hereafter the procedure for a proper turning off of the device is described:<br />

1. Close the software.<br />

2. Shutdown the operating system.<br />

3. Wait for the workstation power off.<br />

4. Turn the unit off by mean of the main switch, which is situated on the Control Box.<br />

Rev. 1.0 - 02/15/2009 Page 8 of 72


About safety<br />

WARNING:<br />

Turn off the device when it has not been used for longer than 3 hours.<br />

WARNING:<br />

Always turn the device off at the end of the working time.<br />

2.3 Emergency shutdown<br />

The device is supplied with two emergency buttons. A first button is placed on the user's table (see figure),<br />

whereas the second one is by the main switch, on the control box. The switching off of the device through an<br />

emergency button causes a prompt break in the x-ray emission and the interdiction of every motor<br />

movement function.<br />

WARNING:<br />

The emergency button must be used only in a danger situation, such as:<br />

• The X-Ray source does not stop emitting;<br />

• Dangerous conditions may damage people, the environment or the device;<br />

• Whenever the system signals an emergency state.<br />

Remote emergency button<br />

Rev. 1.0 - 02/15/2009 Page 9 of 72


About safety<br />

2.4 Patient's and user's safety guidelines<br />

Risks for a patient and for involved people are avoided working in a proper way and correctly positioning the<br />

patient. Special care should be taken in case of disabled or injured people.<br />

2.4.1 Patient positioning<br />

Be sure that a patient is properly placed inside the scan area, with his/her head on the head support, and<br />

that no part of his/her body may impact the device or be pressed during the positioning and the examination.<br />

Be sure that a patient's garments and hair may not get caught. Similarly regarding catheters and ECG<br />

cables.<br />

Do not perform any movement until the patient's safety is assured and there are no obstacles to the device<br />

movement.<br />

Refer to Par. 6.1.2 - "Positioning the patient and starting a new scan" for a proper positioning for a patient.<br />

Rev. 1.0 - 02/15/2009 Page 10 of 72


About safety<br />

2.4.2 Patient's exit<br />

After the end of an exam or after an emergency shutdown, you can let the patient go without waiting for the<br />

rotating arm to return to the start position.<br />

2.4.3 Protection from ionizing radiation<br />

WARNING:<br />

<strong>NewTom</strong> <strong>VG</strong>i is a radiological device and therefore it exposes operators and<br />

patients to the risks consequent to the exposure to ionizing radiation. It must<br />

be used according to safety rules that are contemplated by the local laws<br />

regarding this matter.<br />

Always carefully follow the applicable regulations about radio-protection and the directions from a Certified<br />

Radiation Protection Expert.<br />

Operator<br />

An operator must oversee the examination process from his/her control position according to the laws in<br />

force; NO ONE SHOULD BE BY THE PATIENT DURING THE EXAMINATION PROCESS.<br />

If a patient panics and an operator's intervention is needed during the examination process, the operator<br />

must be equipped with proper protection clothing and devices, according to the laws.<br />

WARNING:<br />

Never stand close to the device during x-ray emission.<br />

Patient<br />

It is the user's responsibility to protect the patients from needless exposures to x-ray.<br />

NOTE:<br />

Consider the possibility to use a leaded cloth to protect the patient from the<br />

scattered radiation.<br />

Devices displaying the emission<br />

The x-ray emission is signalled by:<br />

1. A signal on the workstation screen such as the one reported below. It appears on video just after the<br />

START command has been sent via keyboard or mouse (see Chapter 6 "Scanning"). It is displayed all<br />

along the scan process.<br />

2. Light indicators (LEDs) inside the control panels. They can be found on the sides of the Scanner Unit.<br />

They light on just after the START command has been sent via keyboard or mouse (see Chapter 6<br />

"Scanning"). They stay on all along the scan process.<br />

Rev. 1.0 - 02/15/2009 Page 11 of 72


About safety<br />

WARNING:<br />

If the x-ray emission signals are on when the specific command has not been<br />

sent or if they are off after a START emission command or if the x-ray emission<br />

does not stop after the supposed time, IMMEDIATELY TURN THE DEVICE OFF<br />

and contact technical assistance.<br />

2.5 Laser exposure<br />

WARNING:<br />

Do not stare into the laser beams, do not view directly with optical instruments,<br />

and avoid direct exposure to the beam. The beams can cause permanent eye<br />

damage.<br />

WARNING:<br />

Keep your eyes at least 50 cm from the laser when the beam is on.<br />

If necessary, use appropriate protective glasses.<br />

WARNING:<br />

The use of controls, adjustments, or the performance of procedures other than<br />

those specified herein may result in hazardous radiation exposure.<br />

The device is equipped with a cross shaped positioning laser. The beam is emitted from an aperture in the<br />

front cover.<br />

The vertical line shows the reconstructed volume sagittal central plane. The horizontal line shows the<br />

reconstructed volume axial central plane.<br />

Fori apertura laser<br />

2.6 Safety notes<br />

Rev. 1.0 - 02/15/2009 Page 12 of 72


About safety<br />

<strong>Works</strong>tations, displays, printers, keyboards and other devices which may be connected to the main <strong>NewTom</strong><br />

<strong>VG</strong>i workstation MUST comply with ISO and/or IEC and/or EN standards and/or local laws in force. QR srl is<br />

available for further information.<br />

The manufacturer is not accountable for faults and/or malfunctioning of parts that were not supplied or<br />

approved by QR srl, or not installed by technical staff qualified and certified by QR srl.<br />

Food and beverage may not be placed or eaten/drunk close to the device and the workstation.<br />

Rev. 1.0 - 02/15/2009 Page 13 of 72


Safety and maintenance of the device<br />

3 Safety and maintenance of the device<br />

This chapter includes information about environmental safety and care for the device. General information<br />

and procedures regarding the device maintenance are also provided.<br />

The user is responsible for a proper use of the system, according to the instructions and procedures which<br />

are described inside this manual. Particularly an user must comply with the following points:<br />

• The device may be used only by authorized and properly trained (about the system and the<br />

protection from radiation exposure) personnel. He/she may also know the regulations regarding the<br />

use of radiological devices.<br />

• The device may never be used in case of any evident electrical mechanical or radiological<br />

malfunctioning. Especially a device can not be used whenever the x-ray emission displays and/or<br />

the emergency button are not working properly.<br />

3.1 Installation requirements<br />

The system must be installed in a medical environment, according to a Qualified Expert recommendations.<br />

A device may not be exposed to acids, corrosive substances, saltiness and wet.<br />

Operating temperature: +10° ÷ +35° Celsius.<br />

Operating humidity condition: min 10%, max 85% not condensing.<br />

Installation site minimum dimensions: 80” x 80”.<br />

The device must be installed on a flat horizontal floor.<br />

When using a chair make sure it is a safe supportive chair with a backrest lower than 30”.<br />

The power supply connection must be implemented according to the laws in force and to the instructions<br />

from the Service <strong>Manual</strong>.<br />

Don’t use temporary connections such as adapter and extension cords to connect the computer and<br />

peripherals to the mains.<br />

The equipment should be permanently connected to the mains according to the instructions stated in the<br />

Service <strong>Manual</strong>.<br />

The medical environment for the installation has to be designed by an expert in protection from ionizing<br />

radiation exposure according to the laws in force. Local laws shall also rule the design of the signposting.<br />

WARNING:<br />

Never move a device after it has been installed. Moving a device may damage<br />

people, the device itself or the environment.<br />

Connect only approved peripherals, computer and cables to the equipment as<br />

specified by the manufacturer.<br />

NOTE:<br />

The workstation must be located outside the patient's area (keeping a distance of at<br />

least 1,5m away from the patient) . Cables connected to the workstation can ONLY<br />

be used to connect the workstation itself. Such cable and their connections can<br />

ONLY be handled by authorized and qualified personnel.<br />

Rev. 1.0 - 02/15/2009 Page 14 of 72


Safety and maintenance of the device<br />

3.2 Guidelines for safety<br />

The device is not protected against liquid or gas infiltration. Liquid intrusion may damage the electrical parts<br />

and put the patient, the user and the environment in danger.<br />

Safety systems of the device do not decrease safety measures against the risk of fire in the place in which<br />

the device has been installed.<br />

Electrostatic discharge<br />

Electrostatic discharge may damage electrical parts in the device. Consequently the floor in the installation<br />

room should be made of antistatic materials.<br />

Extinguishers<br />

CO 2 extinguishers should be installed in an easily accessible area.<br />

X-Ray lamp<br />

It is possible for the user to install a x-ray lamp in order to signal the ready and emission states of the x-ray<br />

source.<br />

Door switches<br />

The system contemplates the possibility for the installation of an external switch that can stop the emission<br />

(typically installed on the entrance of the device room).<br />

Electromagnetic compatibility<br />

For information regarding the electromagnetic compatibility please refer to APPENDIX A:<br />

Reference.<br />

Technical<br />

3.3 Changes in the device<br />

Possible changes or updates for the system must comply with the laws in force.<br />

WARNING:<br />

It is forbidden to open or manumit the device with any tool.<br />

Whatever change (hardware or software) that was not p<strong>rev</strong>iously authorized is<br />

forbidden and can inhibit a proper working in the device. It may also damage the<br />

patient, the user, the device and the environment.<br />

Responsibility limits<br />

The manufacturer is not in charge for safety, reliability and performance features in the following<br />

cases:<br />

• Installation, maintenance, changes, repairs and/or updates were not performed by personnel<br />

that was not directly authorized by the manufacturer or distributor.<br />

• Replace parts were not approved by the manufacturer or distributor.<br />

• Environmental conditions were not conform according to the requirements of the laws in force<br />

and to the suggestions of a qualified expert.<br />

• The device is used in a way that is not corresponding to what is described inside this manual.<br />

Rev. 1.0 - 02/15/2009 Page 15 of 72


Safety and maintenance of the device<br />

3.4 Device maintenance<br />

WARNING:<br />

Always turn the power off the device before any maintenance act.<br />

WARNING:<br />

The device has no repairable part inside. Never remove the device covers.<br />

WARNING:<br />

The only part that can be repaired by the user is the input fuse. It is placed in<br />

the input panel, on the control box side. New fuses must comply with the<br />

manufacturer specification.<br />

WARNING:<br />

In order to guarantee protection against the risk of fire, replace only with fuses<br />

of the same type and range.<br />

Regular maintenance<br />

Regular maintenance is required to guarantee a proper working of the device and safety for the patient, the<br />

user and other people.<br />

The device maintenance and repair must be performed only by personnel that was directly authorized by the<br />

manufacturer or the distributor. Every part of the system can only be tested and if needed substituted by<br />

qualified personnel.<br />

WARNING:<br />

If the x-ray source has not been working for longer than three months, please<br />

perform a beam forming procedure (see "Software <strong>Manual</strong>" - Appendix A).<br />

Dangerous cleaning substances<br />

Some cleaning substances should be avoided to p<strong>rev</strong>ent damage to people and to the device (see 3.5.1<br />

"Chemical agents to avoid").<br />

P<strong>rev</strong>entive maintenance<br />

Please periodically check the workstation-scanner, workstation-control box and control box power supply<br />

cables. Also verify the connections between the workstation parts (display, keyboard, mouse, printer)<br />

according to the manufacturer instructions.<br />

Parts and accessories storage<br />

Other parts or accessories must be stored according to their technical specifications.<br />

Malfunctioning<br />

If the system is not working as described inside this manual, please immediately contact the technical<br />

service.<br />

Rev. 1.0 - 02/15/2009 Page 16 of 72


Safety and maintenance of the device<br />

Maintenance agreement<br />

The device should be periodically checked: please contact the manufacturer or distributor to discuss for a<br />

maintenance contract.<br />

System testing check list<br />

The following check list specifies the suggested time spaces for the system controls. For further information<br />

please contact Your local distributor.<br />

Responsible Part Activity Frequency<br />

<strong>User</strong> Global system QA phantom check Weekly<br />

Radioprotection<br />

expert (RPE) Global system Radiological routine In charge of the<br />

RPE<br />

Error Log Check 12 months<br />

Every external part Damage check 12 months<br />

Technical staff Emergency button Emergency test 12 months<br />

Electrical functioning Check 12 months<br />

Mechanical functioning Check 12 months<br />

3.5 Cleaning and disinfection<br />

This chapter gives general information about cleaning and disinfection of the device.<br />

WARNING:<br />

Power the device off before any cleaning.<br />

Some cleaning or disinfecting products may damage Your health. Such products concentration in the air<br />

should never exceed the local laws' limits. In this aim always follow the manufacturer instructions and be<br />

sure that the room is well airy.<br />

3.5.1 Chemical agents to avoid<br />

• Do not use spray or solvent which may enter the device, causing damage electrical parts or forming<br />

inflammable mixtures. This could put in danger the patient, the user and the environment.<br />

• Do not use abrasive liquids such as acetone, gas oil or alcohol (exception: alcohol can be used for the<br />

cleaning of the workstation display). These substances can deteriorate surfaces and materials that are<br />

responsible for the device proper working.<br />

• Do not use substances that may release ammonia due to dissociation or decomposition. Ammonia can<br />

corrode.<br />

• Do not use cleaning products containing silicone: it tends to accumulate and possibly damage electrical<br />

contacts.<br />

Rev. 1.0 - 02/15/2009 Page 17 of 72


Safety and maintenance of the device<br />

3.5.2 Cleaning<br />

3.5.2.1 Scanner and control box<br />

The device must be cleaned with a wet cloth and a soft solution. The cloth should be wet before applying it<br />

on the scanner surface.<br />

WARNING:<br />

Do not apply cleaning solutions directly on the device surface.<br />

3.5.2.2 <strong>Works</strong>tation and periphera ls<br />

Follow the manufacturer's instructions for cleaning the workstation and its peripheral units. If these are<br />

missing, please refer to the p<strong>rev</strong>ious paragraph.<br />

To clean the display use a wet cloth with alcohol or a solution made up of 1/3 to 2/3 distilled water and<br />

alcohol.<br />

NOTE:<br />

For further information about safety and maintenance of the device please contact<br />

your local distributor.<br />

3.5.3 Disinfection<br />

Despite the device do not require any disinfection, it is suggested to use standard disinfection product for the<br />

cleansing of the head support.<br />

3.5.4 Sterilization<br />

No sterilization is required for the normal use of the device.<br />

3.6 Transport and storage<br />

During transport and storage the following conditions must be observed:<br />

Transport and storage temperature:<br />

Transport and storage humidity:<br />

-20° to +70° (Ce lsius)<br />

min 10%, max 85% not condensing.<br />

Do not expose to acids, saltiness, rain.<br />

Rev. 1.0 - 02/15/2009 Page 18 of 72


Safety and maintenance of the device<br />

3.7 Device disposal<br />

3.7.1 Information for the device owner<br />

This symbol, on a device label, shows that the product can not be disposed along with other urban<br />

garbage.<br />

The separate collection of this equipment at the time of disposal is managed by the dealer. When<br />

disposing the equipment the user should contact the dealer and follow the suggested procedure to<br />

allow the separate collection and recycling of this equipment at the time of disposal.<br />

The separate collection and recycling of your waste equipment at the time of disposal will help to<br />

conserve natural resources and ensure that it is recycled in a manner that protects human health<br />

and the environment.<br />

In case of unauthorized disposal of the device there are sanctions that vary according to the local<br />

and/or regional regulations.<br />

For workstations and every other peripheral disposal it is necessary to refer to the attached instructions<br />

provided by their manufacturers.<br />

3.7.2 Information for the Disposing/Recycle centers<br />

Divide the x-ray source, the mechanical and electrical parts, plastic covers and workstation with its<br />

peripherals.<br />

X-ray sources contain oil that must be extracted for its proper disposal.<br />

Plastic parts must be disposed through approved procedures.<br />

For other parts that have no specific instructions by its manufacturer, please refer to the local laws and<br />

guidelines about health, safety and environment protection.<br />

Rev. 1.0 - 02/15/2009 Page 19 of 72


Getting started…<br />

4 Getting started…<br />

This chapter provides a brief introduction for the <strong>NewTom</strong> <strong>VG</strong>i system, its power on and off routines and<br />

control devices.<br />

4.1 An introduction to the system<br />

The device <strong>NewTom</strong> <strong>VG</strong>i is a panoramic 3D system, dedicated to dento-maxillo-facial area imaging. It<br />

performs the so called "cone-beam" technology.<br />

<strong>NewTom</strong> <strong>VG</strong>i is intended for diagnosis of the dento-maxillo-facial complex.<br />

It’s designed for:<br />

<br />

<br />

<br />

imaging of temporo mandibular joints;<br />

imaging of mandibula/jaw for surgical planing;<br />

imaging of nasal/sinus and maxillofacial complex;<br />

A patient is made to lean on the head support and centered by means of 2 laser modules and "scout view"<br />

images.<br />

The scanning system performs a completely rotation around the patient's head. Radiological images are<br />

acquired, that are then automatically processed by the system. The result is the slices set that forms the<br />

reconstructed volume. At the end of this process the axial slices set composes the Volumetric Data. Through<br />

these data it is possible to display coronal and sagittal views of the reconstructed volume in real time.<br />

After defining a Region Of Interest (ROI), from the volumetric data the user can start the creation of a study.<br />

The ROI can be inclined from the volumetric data both to obtain perpendicular images and to correct<br />

positioning errors.<br />

Working on the study data it is possible to create panoramic, transaxial and 3D images. You can also work<br />

on these kinds of images measuring distances, angles, putting comment etc.<br />

Finally new images can be saved inside the study.<br />

The study images can be used to compile a report, that can be then printed and/or saved on a physical<br />

support.<br />

To study in deep these themes please refer to the Software <strong>Manual</strong>.<br />

4.2 Working principle<br />

According to the cone-beam technology, the source detector system performs a single rotation around the<br />

patient's head, simultaneously acquiring every necessary data for the volumetric reconstruction. Data<br />

acquired each scan step are the digital images corresponding to the radiographic projection. The raw data<br />

set so collected is used in the volumetric tomographic reconstruction process.<br />

This technology brings some advantages:<br />

• direct reconstruction of any set of the scanned object points without passing through axial reconstruction<br />

and data re-formatting;<br />

• total scan time related to the acquisition electronics, rather than to the x-ray tube power and the<br />

mechanics, usually shorter;<br />

• under same conditions of total scan time, less requirements in regards to the source/tube assembly<br />

power and scan mechanics, with constructive and maintenance advantages.<br />

Rev. 1.0 - 02/15/2009 Page 20 of 72


Getting started…<br />

4.3 Overall view<br />

The system is made up of three main parts: scanner unit, control box and main workstation. It is also<br />

possible to add more workstations for data processing and storage. This subject is discussed in the<br />

"Software <strong>Manual</strong>".<br />

NOTE:<br />

The system may not be expanded with parts or accessories other than the ones<br />

described inside this manual.<br />

Rev. 1.0 - 02/15/2009 Page 21 of 72


Getting started…<br />

4.4 Scanner unit<br />

4.4.1 Controls placed on the scanner unit<br />

The scanner unit is the main part of the system, it performs the scan. Two control panels are placed on the<br />

columns side. They have the same appearance and function: the control of the positioning movements, the<br />

light display of the x-ray source emitting state and the control of the positioning laser beam.<br />

Hereafter a short description of each button and sign.<br />

X-ray light display:<br />

the LEDs turn on during the x-ray.<br />

Vertical movement enable button:<br />

it enables the functioning of the buttons for vertical movements.<br />

Vertical movement buttons (up/down):<br />

to move up or down the structure supporting the rotating arm, press<br />

one of them simultaneously with the corresponding enabling button.<br />

Rotation movement enable button:<br />

it enables the functioning of the buttons for rotating arm movements.<br />

Rotating arm movement buttons:<br />

to rotate the arm clockwise or anti-clockwise, press one of them<br />

simultaneously with the corresponding enabling button.<br />

Laser button (only for devices with positioning laser beam):<br />

press to turn the positioning laser beam on/off. The laser will stay light<br />

up for about 60 seconds<br />

Rev. 1.0 - 02/15/2009 Page 22 of 72


Getting started…<br />

4.4.2 Controls on the head support keyboard<br />

Now it comes described the function of every button:<br />

Horizontal movement buttons (DX/SX):<br />

allows the movement of the head support system in the horizontal<br />

direction (right / left)<br />

Vertical movement buttons (UP/DOWN):<br />

allows the movement of the head support system in the vertical<br />

direction (up / down)<br />

Transversal movement buttons (FR/RR):<br />

allows the movement of the head support system in the transversal<br />

direction (front / rear)<br />

Rev. 1.0 - 02/15/2009 Page 23 of 72


Getting started…<br />

4.5 Control box<br />

The control box contains the electronics devices that manage the functioning of the system.<br />

4.5.1 Main switch<br />

On the external surface of the control box You can find the main switch to turn the system on/off and the fuse<br />

support.<br />

Main Switch<br />

Emergency Button<br />

Input Fuse Holder<br />

4.5.2 Input panel<br />

On the same side of the main switch You can find the panel from which cables depart to the power supply, to<br />

the scanner and to the CAN bus interface on the workstation.<br />

CAN bus connection to<br />

workstation<br />

Optional output for door switch<br />

Optional output<br />

for external lamp<br />

Power supply input<br />

Output for table<br />

emergency button<br />

Rev. 1.0 - 02/15/2009 Page 24 of 72


Getting started…<br />

4.6 Standard accessories<br />

Along with the device some standard accessories are supplied. They are:<br />

Head support:<br />

it is fixed to the scanner unit. It makes possible for correct<br />

patient's positioning and p<strong>rev</strong>ents excessive movement of<br />

the patient during the scan.<br />

QA phantom:<br />

used for the execution of the quality assurance procedure.<br />

Calibration base:<br />

it allows for placing the QA phantom on the scanner unit.<br />

Rev. 1.0 - 02/15/2009 Page 25 of 72


Getting started…<br />

4.6.1 Cables<br />

The device also includes linking cables between workstation-scanner unit and workstation-control box. They<br />

are:<br />

Ethernet cable (4 pairs/26 AWG-FTP-5E) (workstation scanner unit)<br />

CAN bus cable (2 pairs/24 AWG-shielded) (workstation control box).<br />

Furthermore, a bundle of cables exits the control box and connects to the scanner unit.<br />

The manufacturer also provides the power supply cable with an end connected to the control box, it is in<br />

charge of the user to connect it to the mains during installation.<br />

WARNING:<br />

Accessories, transducer and cables different from the ones here specified may<br />

change the electromagnetic compatibility characteristics of the device.<br />

4.7 Optional accessories<br />

At the present time there are no optional accessories for a <strong>NewTom</strong> <strong>VG</strong>i system.<br />

Rev. 1.0 - 02/15/2009 Page 26 of 72


Getting started…<br />

4.8 System startup<br />

To properly turn the device on:<br />

1. Power the scanner unit on with the main switch (on the control box).<br />

2. Turn on the workstation. (Please refer to its manual).<br />

3. Wait for the workstation operating system to startup.<br />

4. Log in the operating system with user name and password as described in the Software <strong>Manual</strong>.<br />

5. NNT application will automatically open.<br />

NOTE:<br />

Always turn the scanner unit on first. If You try to use the NNT software before the<br />

appliance has done its initialization successfully, an error will occur (see 8.1 -<br />

Communications problem).<br />

4.9 System shutdown<br />

Please refer to the Par. "2.2 - Turning the device off".<br />

Rev. 1.0 - 02/15/2009 Page 27 of 72


Preliminary procedures<br />

5 Preliminary procedures<br />

This chapter describes the procedures to perform compulsorily before starting scanning patients.<br />

These procedures are three:<br />

• X-Ray source conditioning;<br />

• Daily check;<br />

• Blank acquisition;<br />

This chapter describes also some required and/or useful functions:<br />

• Beam limiter test<br />

• Completely lift down the rotating arm<br />

X-Ray source conditioning and Blank acquisition must be performed every two weeks (14 days), while Daily<br />

check is mandatory every day, before patients acquisitions.<br />

If these procedures are not performed the software will disable the scan functions.<br />

Operating modes are described in detail in the same named chapter from the "Software <strong>Manual</strong>".<br />

5.1 X-Ray source conditioning<br />

The conditioning procedure is needed to prepare the x-ray source for its routine working.<br />

The software asks to perform it every two weeks. It is suggested to perform the conditioning also if the<br />

device does not operate for days.<br />

REMOVE<br />

Before starting the procedure, please verify that the<br />

scan area is completely free from objects.<br />

Remove the head support.<br />

At the end of conditioning process, the software will automatically starts Daily check and Blank acquisition<br />

procedures.<br />

5.2 Daily Check<br />

By mean of the Daily Check procedure the system verifies that each every part of itself properly works.<br />

Rev. 1.0 - 02/15/2009 Page 28 of 72


Preliminary procedures<br />

5.3 Blank acquisition<br />

Blank acquisition allows to optimize the scan performance acquiring a background image.<br />

This procedure is request by the software every two weeks.<br />

REMOVE<br />

Before starting the procedure, please verify that the<br />

scan area is completely free from objects.<br />

Remove the head support.<br />

The resulting blank acquisition image will look like<br />

shown in the figure. Please verify that no artifacts or<br />

objects appear.<br />

5.3.1 Invalidate blank acquisition<br />

To invalidate blank acquisition follow the steps below:<br />

Select the wished detector field<br />

Rev. 1.0 - 02/15/2009 Page 29 of 72


Preliminary procedures<br />

From the NNT software main toolbar select Scan <br />

Invalidate Blank.<br />

At the next detector field selection the NNT software<br />

will ask a new blank acquisition<br />

5.4 Beam limiter test<br />

NOTE:<br />

This procedure applies only to devices with multiple FOV.<br />

From the NNT software main toolbar select Tools Scanner Test. The service window will appear.From<br />

the service window toolbar select the Tools Beam Limiter Test. To verify hires field lead collimator<br />

position:<br />

Press “Beam limiter test Acquisition” button.<br />

Verify the acquired image: the light gray rectangle must be vertically center between the two red lines.<br />

5.5 Completely lift down the rotating arm<br />

This function is useful to completely move down the rotating arm. For example during transport , in case of<br />

mobile application , is suggested to position the arm completely down.<br />

WARNING:<br />

Please, verify there are no objects that can obstruct the downward movement of<br />

the arm before using this function!<br />

To completely lift down the rotating arm follow the steps below:<br />

1. From the NNT software main toolbar select Tools Scanner Test. The service window will appear.<br />

2. From the service window toolbar select the Tools Lift down the rotating arm.<br />

Rev. 1.0 - 02/15/2009 Page 30 of 72


Scanning<br />

6 Scanning<br />

This chapter describes the procedures to follow for a proper patient's or denture positioning and examination.<br />

The exact procedure to perform a scan on a patient or on a denture can be found in the corresponding<br />

chapter inside the Software <strong>Manual</strong>.<br />

Please also refer to:<br />

Chapter 2 "About safety";<br />

Chapter 3 "Safety and maintenance of the device".<br />

It is possible to perform the scanning in 3 modalities:<br />

Full field ("Full"), High Resolution Zoom ("HiRes Zoom")<br />

and "Zoom".<br />

Select the modality in the "Scan Manager" panel.<br />

6.1 Scanning a patient<br />

6.1.1 Preparing the patient<br />

An important step in a scan routine is well preparing the patient. This can contribute to a proper scanning<br />

and to high quality images. The aim of such preparing is making the patient relaxed and in a good position<br />

before and during the scan. Here are some suggestions that can help in this concern.<br />

Room setting<br />

Be sure that the scanner unit is clean and ready to scan ("Daily Check" and "Blank Acquisition" already<br />

performed).<br />

Preparing the patient<br />

Ask the patient to remove if present any jewelry (earrings, necklaces), glasses and removable metallic<br />

prothesis.<br />

Accommodate the patient<br />

After to have made accommodate the patient in the scanning area, set the scanner and the head support<br />

system so as to frame the interested area and the neck/bust of the patient assumes an erected position.<br />

Particular patients<br />

A special care has to be taken in case of children, elderly people, claustrophobic people, physically or<br />

mentally disabled.<br />

Explanation of the examination<br />

Please explain in short the scan routine to the patient, including the data collection process, the positioning<br />

and scanning.<br />

Proper breathing<br />

Ask the patient to breathe slowly during the scanning (a slow and continuous breath helps from swallowing).<br />

Rev. 1.0 - 02/15/2009 Page 31 of 72


Scanning<br />

Relaxation<br />

Ask the patient to keep the mouth closed without grinding the teeth together.<br />

Avoiding delays<br />

Good examination times can be obtained also by carrying out all the preliminary procedures before starting<br />

the scan itself.<br />

Voice instructions<br />

Introduce the patient possible voice instructions that the operator may give during the scan.<br />

Rev. 1.0 - 02/15/2009 Page 32 of 72


Scanning<br />

6.1.2 Positioning the patient and starting a new scan<br />

Hereafter is a description of how to position and center a patient inside the scan area. Perform these action<br />

just when the software signals them.<br />

WARNING:<br />

The scan area (the place in which a patient is positioned during the scan) must<br />

be free from any objects, except from the head support. They may damage the<br />

patient or invalidate the scan results.<br />

C<br />

B<br />

A<br />

Set the head support to the default condition (the<br />

positioning of the three rulers (A, B, C) must be set<br />

on the "D" position) by means of the head support<br />

keyboard (make reference to the Par. “4.4.2 -<br />

Controls on the head support keyboard”)<br />

Using the positioning keyboard located on the side of<br />

the scanner (make reference the Par. “4.4.1 -<br />

Controls placed on the scanner unit”), move the<br />

rotating arm structure vertically so that the patient<br />

can easily enter the gantry (the head support should<br />

be approximately at the patient's head height).<br />

Move the patient to the head support without leaning<br />

on the structure.<br />

Turn on the positioning lasers through the positioning<br />

keyboard.<br />

NOTE:<br />

Ask the patient to close the eyes before turn on the<br />

positioning lasers.<br />

The vertical line crosses through the vertical axis of<br />

the scanned volume, and should correspond to the<br />

patient's sagittal plane.<br />

Scanned<br />

Height when in<br />

zoom mode<br />

Upper horizontal<br />

line<br />

Lower horizontal<br />

line<br />

The upper horizontal line is:<br />

- The center of the scanned volume when in "Full"<br />

Volume mode<br />

- The upper limit of the scanned volume when in<br />

one of the "Zoom" mode<br />

The lower horizontal line is the lower limit of the<br />

scanned volume in both "Full" Volume mode and<br />

"Zoom" modes.<br />

Rev. 1.0 - 02/15/2009 Page 33 of 72


Scanning<br />

Set the height of scanner with the positioning<br />

keyboard so to frame the area of interest with the<br />

laser lines (the lower line being the lower limit of the<br />

scanned volume).<br />

When in "Zoom" mode, adjust the height of the chin<br />

support with the head support keyboard, so to level<br />

with the patient's chin.<br />

Make the patient lean the head onto the chin and the<br />

head support, and ask him/her to grab the handle at<br />

the sides.<br />

You may now want to finely further adjust the location<br />

of the scanned volume with the head support<br />

keyboard.<br />

NOTES:<br />

Head support at the highest position is for<br />

mandibulary "Zoom" mode, at the lowest position is<br />

for maxillary "Zoom" mode.<br />

If the patient ends up in a forced position, you may<br />

need to re-adjust the height of the gantry via the<br />

positioning keyboard.<br />

Stabilize the patient’s head with the provided strap.<br />

After to make sure that the rotating arm does not hit<br />

the patient's shoulders (rotate shortly the arm through<br />

the movements keyboards), proceed with the<br />

verification of the corrected positioning of the patient<br />

by means of the AP (anterior - posterior) and the LL.<br />

Rev. 1.0 - 02/15/2009 Page 34 of 72


Scanning<br />

6.2 Scanning a denture<br />

6.2.1 Preliminary operations<br />

Data about the denture are to be inserted, if they are not already in the software database. For details refer<br />

to the Software <strong>Manual</strong>.<br />

6.2.2 Denture positioning<br />

Hereafter are the operations to perform to properly position and center the denture inside the scanning area.<br />

Perform such steps according to the software instructions.<br />

Remove from the head support the movable lower<br />

part for the patient's chin.<br />

Fix the denture to a proper support in such a way<br />

that it assumes the same position as in the patient's<br />

mouth.<br />

A better centering of a patient's head is achievable<br />

through the use of the AP (anterior - posterior) and<br />

the LL (lateral - lateral) scout views.<br />

Then it is possible to perform the scanning, as<br />

described in the "Software <strong>Manual</strong>".<br />

Rev. 1.0 - 02/15/2009 Page 35 of 72


Quality assurance<br />

7 Quality assurance<br />

The quality assurance process consists in performing a routine scan on the specific phantom, through an<br />

automated procedure. It is recommended to perform such a test at least once a week, in order to assure the<br />

proper functioning of the device and the validity of a scan results.<br />

Before starting the Quality Assurance (QA) procedure the FOV must be selected.<br />

The test procedure is described inside the Software <strong>Manual</strong>.<br />

7.1 Phantom positioning<br />

Hereafter the procedure to properly place and center the QA phantom inside the scan area is described.<br />

Please perform these actions exactly when the software signal them.<br />

REMOVE<br />

Remove the head support.<br />

Fix the calibration base.<br />

Rev. 1.0 - 02/15/2009 Page 36 of 72


Quality assurance<br />

B<br />

A<br />

C<br />

Place the QA phantom on its base in a central<br />

position.<br />

In order to facilitate the correct positioning of the<br />

phantom, set up the head support being used the<br />

keys of the head support keyboard (make reference<br />

to the Par. “4.4.2 - Controls on the head support<br />

keyboard”).<br />

Ruler A (Height) position “QA”<br />

Rulers B and C (width, depth) default position (“D”)<br />

1a scout view<br />

The first displayed scout view represents the<br />

phantom as seen by the scanner in the start position.<br />

Using this reference one can center the phantom in<br />

the direction perpendicular to the rotating arm.<br />

Repeat this step as many times as necessary (the<br />

aluminium cylinder should appear between the<br />

vertical sketched red lines and the inner small<br />

spheres to the QA should be crossed from the<br />

horizontal sketched red lines).<br />

Frontally center the QA phantom using the<br />

information from the second scout view, which is<br />

acquired as the rotating arm is in a 90 degrees<br />

position.<br />

It is now possible to perform the scan of the<br />

phantom.<br />

For more details about the QA scan please refer to<br />

the "Software <strong>Manual</strong>".<br />

2a scout view<br />

Rev. 1.0 - 02/15/2009 Page 37 of 72


Quality assurance<br />

7.2 Sample <strong>Image</strong>s<br />

Hereafter some sample images from a QA phantom analysis.<br />

Side view.<br />

Axial view.<br />

Panoramic view.<br />

7.3 Storage of QA data<br />

The software automatically stores each phantom analysis report. The reports can be opened by selecting<br />

from the main menu View QA Report.<br />

To move among the reports list use the PAGE DOWN, PAGE UP keyboard button.<br />

It is possible to save a copy of the QA report analysis in PDF format: select menu File Save as PDF.<br />

It is recommended to save a printed copy of each report.<br />

Rev. 1.0 - 02/15/2009 Page 38 of 72


Troubleshooting<br />

8 Troubleshooting<br />

The following tables list problems that may occur and quick instructions to follow to solve them. If the "What<br />

to do…" column lists more than a solution, please try with the first one and so on till the problem full<br />

resolution.<br />

8.1 Communications problem<br />

The workstation cannot connect to the scanner unit or to the control box<br />

ErrCode Hal Error Cause What to do…<br />

2103 n.a. Emergency button is pressed. Unlock the emergency button.<br />

17601 16-001<br />

The device is not powered.<br />

The CANbus cable might be<br />

disconnected.<br />

See 4.8 System startup<br />

Connect and/or verify the cable.<br />

17601 20-001<br />

The Ethernet cable might be<br />

disconnected.<br />

Connect and/or verify the cable.<br />

8.2 Rotating arm<br />

The rotating arm does not move<br />

Cause<br />

What to do…<br />

The internal part controlling the rotating arm Close the software, turn the device off, wait for about ten seconds,<br />

motor encountered a problem.<br />

then retry. See 4.8 System startup.<br />

8.3 <strong>Image</strong> acquisition<br />

<strong>Image</strong> quality<br />

Description Probable cause What to do…<br />

Reconstructed axial<br />

images are completely<br />

dark.<br />

Blank acquisition may have been<br />

performed with some object or even<br />

someone inside the scan area.<br />

Perform a new Blank Acquisition. The scans that<br />

resulted in dark images must be repeated.<br />

Every reconstructed axial<br />

images contain white<br />

circular artifacts.<br />

The head support structure was<br />

mounted during the Blank Acquisition<br />

routine.<br />

Perform a new Blank Acquisition. The scans that<br />

resulted in dark images must be repeated.<br />

Rev. 1.0 - 02/15/2009 Page 39 of 72


IEC61223: Acceptance Test<br />

9 IEC61223: Acceptance Test<br />

A description of the software utilities for the execution of acceptance test can be found in the Software<br />

<strong>Manual</strong>. Here are reported specific instructions to perform such tests (please refer to norm IEC61223-3-5)<br />

along with the corresponding paragraph from the norm.<br />

9.1 Positioning of the patient support [paragraph 5.1]<br />

Not applicable.<br />

9.2 Patient positioning accuracy [paragraph 5.2]<br />

The <strong>NewTom</strong> <strong>VG</strong>i (version Zoom/HiRes) is equipped with cross-shaped laser pointer aimed to help the<br />

patient positioning. The horizontal line locates the scanned volume central axial plane, the vertical line<br />

locates the scanned volume central sagittal plane. No location pointer has been provided for coronal plane<br />

location.<br />

9.2.1 Axial patient positioning accuracy [paragraph 5.2.1]<br />

Position an object along the axial plane located by the horizontal line and scan it.<br />

See Software manual for data analysis.<br />

9.2.2 Sagittal and coronal patient positioning light [paragraph 5.2.2]<br />

Position an object along the sagittal plane located by the horizontal line and scan it.<br />

See Software manual for data analysis.<br />

9.3 Tomographic section thickness [paragraph 5.3]<br />

9.3.1 Tomographic section thickness for axial scanning [paragraph 5.3.1]<br />

See "Software <strong>Manual</strong>".<br />

9.3.2 Tomographic section thickness for helical scanning [paragraph 5.3.2]<br />

Not applicable.<br />

9.4 Dose [paragraph 5.4]<br />

Take required scans for each modality according to IEC 60601-2-44. The machine calculates automatically<br />

the loading factors, according to the scanned object. There is no way to select manually loading factors for a<br />

scan.<br />

9.5 Noise, mean CT number and uniformity [paragraph 5.5]<br />

See "Software <strong>Manual</strong>".<br />

9.6 Spatial resolution [paragraph 5.6]<br />

See "Software <strong>Manual</strong>".<br />

Rev. 1.0 - 02/15/2009 Page 40 of 72


IEC61223: Acceptance Test<br />

9.7 IEC 61223-3-5 reference table<br />

This table supplies reference values and reference conditions to be used when applying the IEC 61223-3-5<br />

as acceptance test, when specified by contractual agreements.<br />

5.1 Positioning of the<br />

patient support<br />

N.A.<br />

5.2 Patient positioning<br />

accuracy<br />

5.2.1 Axial patient positioning<br />

accuracy<br />

±10mm from isocenter<br />

5.3 Tomographic section<br />

thickness<br />

5.2.2 Sagittal and coronal patient<br />

positioning light accuracy<br />

5.3.1 Tomographic section<br />

thickness for axial scanning<br />

5.3.2 Tomographic section<br />

thickness for helical scanning<br />

±10mm from isocenter<br />

Loading factors: automatically selected by the<br />

software.<br />

Thickness<br />

Raccomanded criteria<br />

5.0 mm ±1.0 mm<br />

1.0 mm ±0.5 mm<br />

0.5 mm -0 +0.5 mm<br />

N.A.<br />

5.4 Dose<br />

Loading factors: automatically selected by the software.<br />

Dose Mode: Standart Dose<br />

CTDI FREE AIR<br />

CTDIvol<br />

Full Mode CTDI 100w<br />

Zoom Mode CTDI 100w<br />

HiRes Zoom Mode CTDI 100w<br />

N.A.<br />

Equal to CTDI 100w<br />

2.4 mGy ±30%<br />

2.0 mGy ±30%<br />

7.3 mGy ±30%<br />

Loading factors: automatically selected by the software.<br />

Dose Mode: Boosted Dose<br />

CTDI FREE AIR<br />

CTDIvol<br />

Full Mode CTDI 100w<br />

Zoom Mode CTDI 100w<br />

N.A.<br />

Equal to CTDI 100w<br />

4.5 mGy ±30%<br />

3.3 mGy ±30%<br />

5.5 Noise, mean CT<br />

number and uniformity<br />

Loading factors: automatically selected by the software.<br />

Values are calculated as Arbitrary Unit (a.u.) because of the machine is not designed to<br />

give HU.<br />

Measured with Catphan 500.<br />

Dose Mode: Standart Dose<br />

Standard deviation (water)<br />

Full Mode 340 a.u. ± 80<br />

Zoom Mode 340 a.u. ± 80<br />

HiRes Zoom Mode 400 a.u. ± 80<br />

Rev. 1.0 - 02/15/2009 Page 41 of 72


IEC61223: Acceptance Test<br />

Dose Mode: Boosted Dose<br />

Full Mode 250 a.u. ± 80<br />

Zoom Mode 240 a.u. ± 80<br />

Mean CT number (water)<br />

Full Mode 8000 a.u. ± 1000<br />

Zoom Mode 8000 a.u. ± 1000<br />

HiRes Zoom Mode 8000 a.u. ± 1000<br />

Uniformity (water)<br />

Full Mode less than a 1600 a.u.<br />

Zoom Mode less than a 1100 a.u.<br />

HiRes Zoom Mode less than a 1100 a.u.<br />

5.6 Spatial resolution<br />

Loading factors: automatically selected by the software.<br />

FP MTF 50 (lp/cm) MTF 10 (lp/cm)<br />

(Full Mode) >7 >12<br />

(Zoom Mode) >7 >13<br />

(HiRes Zoom Mode) >9 >17<br />

Rev. 1.0 - 02/15/2009 Page 42 of 72


APPENDIX A: Technical Reference<br />

APPENDIX A: Technical Reference<br />

Rev. 1.0 - 02/15/2009 Page 43 of 72


APPENDIX A: Technical Reference<br />

Scanner unit<br />

Scanning system<br />

(cone beam technology)<br />

Single rotation and volumetric acquisition.<br />

Scan parameters<br />

Full mode:<br />

Time / X-Ray emission time<br />

18 s / 3.6 s<br />

HiRes Zoom Mode:<br />

Time / X-Ray emission time<br />

27 s / 3.6 – 5.4 s<br />

Zoom Mode:<br />

Time / X-Ray emission time<br />

18 s / 3.6 s<br />

Sampling angle 360°<br />

Patient's centring Fixed position Motorized Head support and<br />

positioning lasers<br />

Anatomical analyzed volume Cylinder Full mode:<br />

Ø max 15 cm<br />

h max 15 cm<br />

HiRes Zoom e Zoom:<br />

Ø max 12 cm<br />

h max 7.5 cm<br />

Weight and dimensions<br />

Scanner Unit<br />

Width (diameter occupied<br />

during a scan)<br />

1130 mm<br />

Depth (max)<br />

1500 mm<br />

Height<br />

2300 mm<br />

1970 mm (<strong>VG</strong> Short version)<br />

Max Weight<br />

272 Kg<br />

Control Box<br />

Width<br />

605 mm<br />

Depth<br />

310 mm<br />

Height<br />

925 mm<br />

Weight<br />

100 Kg<br />

Rev. 1.0 - 02/15/2009 Page 44 of 72


APPENDIX A: Technical Reference<br />

Detector<br />

Pixels 1920 x 1536<br />

Pixel size 127 x 127 µm<br />

Pixel depth 14 bit<br />

S/N 90 dB<br />

Frame rate Max 30 F/s<br />

Scout view radiological images<br />

Full mode:<br />

<strong>Image</strong> pixels 960 x 768 Pixels<br />

Pixel depth 14 n.<br />

Pixel Size 0.254 x 0.254 mm<br />

HiRes Zoom mode:<br />

<strong>Image</strong> pixels 1344 x 1088 Pixels<br />

Pixel depth 14 n.<br />

Pixel Size 0.127 x 0.127 mm<br />

Zoom mode:<br />

<strong>Image</strong> pixels 672 x 544 Pixels<br />

Pixel depth 14 n.<br />

Pixel Size 0.254 x 0.254 mm<br />

Reconstructed volume<br />

Full mode:<br />

Shape<br />

Cylinder<br />

Diameter 150 mm<br />

Voxel Size 0.3 mm 3<br />

<strong>Image</strong> pixels 512 x 512 Pixels<br />

Pixel depth 16 bit<br />

HiRes zoom mode:<br />

Shape<br />

Cylinder<br />

Diameter 123 mm<br />

Voxel Size 0.15 mm 3<br />

<strong>Image</strong> pixels 820 x 820 Pixels<br />

Pixel depth 16 bit<br />

Rev. 1.0 - 02/15/2009 Page 45 of 72


APPENDIX A: Technical Reference<br />

Zoom mode:<br />

Shape<br />

Cylinder<br />

Diameter 123 mm<br />

Voxel Size 0.24 mm 3<br />

<strong>Image</strong> pixels 512 x 512 Pixels<br />

Pixel depth 16 bit<br />

Rev. 1.0 - 02/15/2009 Page 46 of 72


APPENDIX A: Technical Reference<br />

Radiological parameters<br />

TUBE IAE mod. x22 0.3/0.6<br />

Rev. 1.0 - 02/15/2009 Page 47 of 72


APPENDIX A: Technical Reference<br />

Rev. 1.0 - 02/15/2009 Page 48 of 72


APPENDIX A: Technical Reference<br />

Rev. 1.0 - 02/15/2009 Page 49 of 72


APPENDIX A: Technical Reference<br />

Rev. 1.0 - 02/15/2009 Page 50 of 72


APPENDIX A: Technical Reference<br />

Rev. 1.0 - 02/15/2009 Page 51 of 72


APPENDIX A: Technical Reference<br />

X-ray tube head<br />

Brand<br />

Modello<br />

P.S.M. s.r.l.<br />

HF1R<br />

X-ray tube IAE X22 0.3/0.6<br />

Classification (IEC 601-1)<br />

Class I Type B<br />

Mechanical data<br />

Head material<br />

Heat storage capacity<br />

Maximum continuos heat dissipation<br />

Aluminum<br />

550 kJ<br />

60 W @ 110kV, 3.6 mA, 10 ms, 15 FPS<br />

Maximum temperature 3 60°<br />

Minimum inherent filtration @ 70 kV<br />

Oil volume compensation<br />

1.4 mm Al<br />

Rubber tank 410 cm³<br />

Dimensions 325 x 145 x 215<br />

Weight<br />

15 kg<br />

Electrical Data<br />

Maximum output voltage<br />

Cathode to Ground<br />

Anode to Ground<br />

Maximum anodic current @ 110 kV<br />

Maximum tube voltage @ 20 mA<br />

Maximum electrical output power 4<br />

Nominal electrical power, 4s emission 5<br />

120 kV<br />

60 kV<br />

60 kV<br />

20 mA<br />

110 kV<br />

2.2 kW<br />

2.2 kW<br />

Maximum power ripple


APPENDIX A: Technical Reference<br />

Cooling Curve<br />

Rotor<br />

HF1R<br />

Startup 230Vac / 0.8s / 10A<br />

Running 60Vac / 2A<br />

X-ray source assembly<br />

X-ray tube head<br />

HF1R<br />

X-ray tube IAE X22 0.3/0.6<br />

Focal spot to image detector distance<br />

630 mm<br />

Focal spot to skin distance (minimum)<br />

150 mm<br />

Total filtration<br />

1.4 mm Al (inherent filtration) + 13 mm Al (Additional filtration)<br />

Beam size (cone aperture)<br />

238 mm x 179 mm (detector area)<br />

X-ray voltage accuracy 6 < 10%<br />

X-ray current accuracy 7 < 20%<br />

Radiation Linearity 8 < 20%<br />

X-Ray emission time accuracy 9<br />

< 10% + 1 ms<br />

INVERTER<br />

Manufacturer<br />

P.S.M. s.r.l.<br />

Model<br />

HF1 3.5kW<br />

6 According to IEC 60601-2-7, par. 50.104.1<br />

7 According to IEC 60601-2-7, par. 50.104.2<br />

8 According to IEC 60601-2-7, par. 50.105.4<br />

9 According to IEC 60601-2-7, par. 50.104.3<br />

Rev. 1.0 - 02/15/2009 Page 53 of 72


APPENDIX A: Technical Reference<br />

INPUT<br />

Maximum power<br />

3.5 kW<br />

Power supply 230 V∼ (± 10%)<br />

Wave form<br />

Sinusoidal 50/60 Hz<br />

Maximum current<br />

16 A<br />

Apparent resistance of supply mains<br />

0.14 ohm<br />

OUTPUT<br />

Peak voltage<br />

350 Vpk<br />

Maximum peak current<br />

120 Apk Max.<br />

Wave form<br />

Sinusoidal 20 kHz<br />

MECHANICAL DATA<br />

Dimensions<br />

160 x 280 x 235 mm<br />

Weight<br />

7 kg<br />

Rev. 1.0 - 02/15/2009 Page 54 of 72


APPENDIX A: Technical Reference<br />

Dose declaration 10<br />

Dose declaration for Standard Dose Mode<br />

Values are expressed in mGy/scan. Their tolerance is ± 30%<br />

FOV Full Mode 11 Zoom Mode 12 HiRes Zoom Mode 13<br />

CTDI 100 Center 2.34 2.10 7.54<br />

CDTI 100 0 2.24 1.80 7.35<br />

CDTI 100 90 2.26 2.00 7.25<br />

CDTI 100 180 2.24 1.80 7.35<br />

CDTI 100 270 2.26 2.00 7.25<br />

CTDI 100 (periferico) 2.25 1.90 7.30<br />

CTDI w 2.34 1.96 7.23<br />

Dose declaration for Boosted Dose Mode<br />

Values are expressed in mGy/scan. Their tolerance is ± 30%<br />

Date: 2009-02-04<br />

FOV Full Mode 14 Zoom Mode 15<br />

CTDI 100 Center 4.48 3.36<br />

CDTI 100 0 4.51 3.21<br />

CDTI 100 90 4.53 3.23<br />

CDTI 100 180 4.51 3.21<br />

CDTI 100 270 4.53 3.23<br />

CTDI 100 (periferic) 4.52 3.22<br />

CTDI w 4.48 3.25<br />

Date: 2009-02-04<br />

10 Measured using “head phantom” according to IEC 60.601-2-44 Sec. 29.1.102.2<br />

11 Loading factors: automaticamente impostati dal software, normalizzati a 110kV, 2.4mA, 3.6s, 8.45mAs<br />

12 Loading factors: automaticamente impostati dal software, normalizzati a 110kV, 3.2mA, 3.6s, 11.34mAs<br />

13 Loading factors: automaticamente impostati dal software, normalizzati a 110kV, 10.9mA, 5.4s, 58.59mAs<br />

14 Loading factors: automaticamente impostati dal software, normalizzati a 110kV, 6.5mA, 3,6s, 23.58mAs<br />

15 Loading factors: automaticamente impostati dal software, normalizzati a 110kV, 6.3mA, 3.6s, 22.86mAs<br />

Rev. 1.0 - 02/15/2009 Page 55 of 72


APPENDIX A: Technical Reference<br />

16<br />

16 Measured using head phantom according to IEC 60601-2-44 par.29.1.102.2<br />

Rev. 1.0 - 02/15/2009 Page 56 of 72


APPENDIX A: Technical Reference<br />

Laser<br />

Laser diode output power<br />

0.9 mW<br />

Wavelength<br />

635 nm<br />

Beam divergence 70º<br />

Pulse length<br />

Continuous wave<br />

Classification Class 1<br />

Other data<br />

Absorbed power<br />

200 V ∼ (± 10%) / 215 V ∼ (± 10%) / 230 V ∼ (± 10%) / 240 V ∼ (± 10%)<br />

50/60 Hz (± 1%)<br />

5 A (during x-ray emission)<br />

0.66 A (stand by)<br />

100 V ∼ (± 10%) / 115 V ∼ (± 10%)<br />

50/60 Hz (± 1%)<br />

10 A (during x-ray emission)<br />

1.16 A (stand by)<br />

Operating temperature: +10 ÷ +35 °C<br />

Operating humidity:<br />

10% ÷ 85 % (not condensing)<br />

Transport and storage temperature: -20 ÷ +70 °C<br />

Transport and storage humidity:<br />

10% ÷ 85 % (not condensing)<br />

Rev. 1.0 - 02/15/2009 Page 57 of 72


APPENDIX A: Technical Reference<br />

Electromagnetic Compatibility<br />

Clause 6.8.3.201 Technical Description - Tab 201 Guidance and manufacturer's declaration -<br />

electromagnetic emissions - for all equipment and systems<br />

TABLE: Guidance and manufacturer's declaration - electromagnetic emissions<br />

The device <strong>NewTom</strong> <strong>VG</strong>i is intended for use in the electromagnetic environment specified below. The customer<br />

or the user of the device <strong>NewTom</strong> <strong>VG</strong>i should assure that is used in such an environment.<br />

Emission Test Compliance Electromagnetic environment – guidance<br />

RF emissions<br />

CISPR11 Group 1 The equipment <strong>NewTom</strong> <strong>VG</strong>i uses RF energy only for its internal function.<br />

Therefore its RF emissions are very low and are not likely to cause any interference<br />

in nearby electronic equipment.<br />

RF emissions<br />

CISPR11<br />

Class A<br />

The equipment <strong>NewTom</strong> <strong>VG</strong>i is suitable for use in all establishments other than<br />

domestic and those directly connected to the public low-voltage power-supply<br />

network that supplies buildings used for domestic purposes.<br />

Harmonic emissions<br />

IEC 61000-3-2<br />

Class A<br />

Voltage fluctuation/flicker<br />

emissions<br />

IEC 61000-3-3<br />

Complies<br />

Rev. 1.0 - 02/15/2009 Page 58 of 72


APPENDIX A: Technical Reference<br />

Clause 6.8.3.201 Technical Description - Tab 202 Guidance and manufacturer's declaration -<br />

electromagnetic immunity - for all equipment and systems<br />

TABLE: Guidance and manufacturer’s declaration – electromagnetic immunity<br />

The equipment <strong>NewTom</strong> <strong>VG</strong>i is intended for use in the electromagnetic environment specified below. The<br />

customer or the user of the device <strong>NewTom</strong> <strong>VG</strong>i should assure that is used in such an environment.<br />

Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment –<br />

guidance<br />

Electrostatic<br />

discharge<br />

(ESD)<br />

IEC 61000-4-2<br />

± 6 kV contact<br />

±8 kV air<br />

6 kV contact<br />

8 kV air<br />

Floors should be wood, concrete, or ceramic<br />

tile.<br />

If floors are covered with synthetic material the<br />

relative humidity should be at least 50 percent.<br />

Electrical fast<br />

transient / burst<br />

IEC 61000-4-4<br />

±2 kV for power supply lines<br />

±1 kV for input/output lines<br />

2 kV<br />

1 kV<br />

Mains power quality should be that of a typical<br />

commercial and/or hospital environment.<br />

Surge<br />

IEC 61000-4-5<br />

±1 kV differential mode<br />

±2 kV common mode<br />

1kV<br />

2kV<br />

Mains power quality should be that of a typical<br />

commercial and/or hospital environment.<br />

Voltage dips, short<br />

interruptions and<br />

voltage variations on<br />

power supply input<br />

lines<br />

IEC 61000-4-11<br />

95% dip in U t) for 0,5 cycles<br />

40% U t<br />

(60% dip in U t) for 5 cycles<br />

70% U t<br />

(30% dip in U t) for 25 cycles<br />

Complies<br />

Complies<br />

Complies<br />

Mains power quality should be that of a typical<br />

commercial and/or hospital environment. If the<br />

user of <strong>NewTom</strong> <strong>VG</strong>i requires continued<br />

operation during power main interruptions, it is<br />

recommended that <strong>NewTom</strong> <strong>VG</strong>i be powered<br />

from an uninterruptible power supply or a<br />

battery.<br />

95% dip in U t) for 5 sec<br />

Functions Interruption<br />

Power frequency<br />

50/60Hz magnetic field<br />

IEC 61000-4-8<br />

3 A/m Complies<br />

Power frequency magnetic fields should be at<br />

least characteristic of a typical location in a<br />

typical commercial or hospital environment.<br />

Rev. 1.0 - 02/15/2009 Page 59 of 72


APPENDIX A: Technical Reference<br />

Clause 6.8.3.201 Technical Description - Tab 204 Guidance and manufacturer's declaration -<br />

electromagnetic immunity - for equipment and systems that are not life-supporting<br />

TABLE: Guidance and manufacturer’s declaration – electromagnetic immunity<br />

The equipment <strong>NewTom</strong> <strong>VG</strong>i is intended for use in the electromagnetic environment specified below. The<br />

customer or the user of the device <strong>NewTom</strong> <strong>VG</strong>i should assure that is used in such an environment.<br />

Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment – guidance<br />

Portable and mobile RF communications equipment<br />

should be used no closer to any part of the <strong>NewTom</strong><br />

<strong>VG</strong>i, including cables, than the recommended<br />

separation distance calculated from the equation<br />

applicable to the frequency of the transmitter.<br />

Conducted RF<br />

IEC 61000-4-6<br />

3 Vrms<br />

150 kHz to 80 MHz<br />

3 Vrms<br />

Recommended separation distance:<br />

d = 1.2 √P<br />

Radiated RF<br />

IEC 61000-4-3<br />

3 V/m<br />

80 MHz to 2,5 GHz<br />

3 V/m<br />

d = 1.2 √P 80 MHz to 800 MHz<br />

d = 2.3 √P 800 MHz to 2.5 GHz<br />

where P is the maximum output power rating of the<br />

transmitter in watts (W) according to the transmitter<br />

manufacturer and d is the recommended separation<br />

distance in meters(m).<br />

Fields strengths from fixed RF transmitters, as<br />

determined by an electromagnetic site survey*, should<br />

be less than the compliance level in each frequency<br />

range**.<br />

Interference may occur in the vicinity of equipment<br />

marked with the following symbol:<br />

Notes:<br />

(1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies<br />

(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,<br />

objects and people<br />

*Field strengths from fixed transmitters, such as base stations for radio (celular/cordless), telephones and land mobile radios, amateur radio,<br />

AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment<br />

due to fixed RF transmitters, an electromagnetic site survey should be performed. If the measured field strength in the location in which the<br />

<strong>NewTom</strong> <strong>VG</strong>i is used exceeds the RF compliance level above, the <strong>NewTom</strong> <strong>VG</strong>i should be observed to verify normal operation. If abnormal<br />

performance is observed, additional measures may be necessary, such as re-orienting or relocating <strong>NewTom</strong> <strong>VG</strong>i<br />

**Over the frequency range 150 kHz to 80 MHz, field strhgths should be less than 3 V/m<br />

Rev. 1.0 - 02/15/2009 Page 60 of 72


APPENDIX A: Technical Reference<br />

Clause<br />

6.8.3.201 Technical Description - Tab 206 Recommended separation distances between<br />

portable and mobile RF communications equipment and the equipment or system - for<br />

equipment and systems that are not life-supporting<br />

TABLE: Recommended separation distances between portable and mobile RF<br />

communications equipment and the equipment<br />

The device <strong>NewTom</strong> <strong>VG</strong>i is intended for use in the electromagnetic environment in which radiated RF<br />

disturbances are controlled. The customer or the user of the device <strong>NewTom</strong> <strong>VG</strong>i can help p<strong>rev</strong>ent<br />

electromagnetic interference by maintaining a minimum distance between portable and mobile RF<br />

communications equipment and the device <strong>NewTom</strong> <strong>VG</strong>i as recommended below, according to the maximum<br />

output power of the communication equipment.<br />

Rated maximum output power of<br />

transmitter<br />

W<br />

Separation distance according to frequency of transmitter<br />

150 kHz to 80 MHz<br />

d=<br />

m<br />

80 MHz to 800 MHz<br />

0,001 0.037 0.037 0.072<br />

d=<br />

800 MHz to 2,5 GHz<br />

d=<br />

0,1 0.37 0.37 0.72<br />

1 1.2 1.2 2.3<br />

10 37.9 37.9 7.27<br />

100 120 120 23<br />

For transmitter rated at a maximum output power not listed above, the recommended separation distance in meters (m) can be estimated<br />

using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts<br />

(w) according to the transmitter manufacturer.<br />

Notes:<br />

(1) At 80 MHz and 800MHz, the separation distance for the higher frequency range applies<br />

(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,<br />

objects and people.<br />

Rev. 1.0 - 02/15/2009 Page 61 of 72


APPENDIX B: Standards conformity<br />

All components, accessories, spare parts must be approved and supplied by QR S.r.l.<br />

Particularly, the connection cables must be ONLY the ones described under 4.6.1 “Cables”.<br />

WARNING:<br />

Use of accessories, transducers, and cables other than those specified may<br />

result in degraded electromagnetic compatibility performance of this device!<br />

CAUTION!<br />

<strong>NewTom</strong> <strong>VG</strong>i should not be stacked directly on top of other equipment, and<br />

other equipment should not be stacked on top of <strong>NewTom</strong> <strong>VG</strong>i. If stacking is<br />

necessary, observe <strong>NewTom</strong> <strong>VG</strong>i to verify normal operation in the stacked<br />

configuration in which it will be used!<br />

Essential performance<br />

In case of a scan stopped by a temporary or permanent malfunctioning, the operator will have the<br />

opportunity to store the data acquired until the failure appeared.<br />

The quality of the reconstructed images will vary depending on the size of acquired data but it will be<br />

probably lower compared to the standard performance of the equipment.<br />

Rev. 1.0 - 02/15/2009 Page 62 of 72


APPENDIX B: Standards conformity<br />

APPENDIX B: Standards conformity<br />

Rev. 1.0 - 02/15/2009 Page 63 of 72


APPENDIX B: Standards conformity<br />

The <strong>NewTom</strong> <strong>VG</strong>i equipment has been built in conformity with IEC normative regarding safety of electricmedical<br />

devices of similar typology and, particularly, with the normative:<br />

• IEC 60601-1: 1988-12 Medical electrical equipment - Part 1: General requirements for safety<br />

• IEC 60601-1-1: Edition 2.0 2000-12 Medical electrical equipment - Part 1-1: General requirements for<br />

safety - Collateral standard: Safety requirements for medical electrical systems<br />

• IEC 60601-1-2: 2001-09 Medical electrical equipment - Part 1-2: General requirements for safety -<br />

Collateral standard: Electromagnetic compatibility - Requirements and tests<br />

• IEC 60601-1-3:1994-07 Part 1: General requirements for safety - 3. Collateral standard: General<br />

requirements for radiation protection in diagnostic X-ray equipment<br />

• IEC 601-1-4: Edition 1.1 2000-04 Medical electrical equipment - Part 1-4: General requirements for safety -<br />

Collateral Standard: Programmable electrical medical systems<br />

• IEC 601-2-7: Edition 2 1998-02 Medical electrical equipment - Part 2-7: Particular requirements for the<br />

safety of high-voltage generators of diagnostic X-ray generators<br />

• IEC 601-2-28: Edition 1.0 1993-03 Medical electrical equipment - Part 2: Particular requirements for the<br />

safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis<br />

• IEC 601-2-32: Edition 1.0 1994-03 Medical electrical equipment - Part 2: Particular requirements for the<br />

safety of associated equipment of X-ray equipment<br />

• IEC 601-2-44: Edition 2.1, 2002-11 Medical Electrical Equipment - Part 2.44:Particular requirements for the<br />

safety of X-ray equipment for computed tomography<br />

• IEC 60825-1: 2003-02 Edition 4, Safety of laser products - Part 1: Equipment classification, requirements<br />

and user's guide<br />

• 21 CFR 1020.33 Computed tomography (CT) equipment April 1, 2006]<br />

CLASSIFICATION IEC 601-1<br />

Protection against electrical shock<br />

Degree of protection against electric shock<br />

Use with flammable anesthetics<br />

CLASS I<br />

TYPE B<br />

Have not been evaluated for use in the<br />

presence of a flammable anaesthetic mixture<br />

with air, oxygen, or nitrous oxide.<br />

Sterilization and disinfection methods<br />

The device is supplied not sterile and it must<br />

not be subjected to sterilization.<br />

See chapter 3.5 “Cleaning and disinfection”<br />

Mode of operation<br />

Continuous with intermittent loading.<br />

Duty cycle<br />

18sec ON / 15 min. OFF<br />

modalities<br />

Full and Zoom<br />

27sec ON / 15 min. OFF HiRes Zoom modality<br />

Rev. 1.0 - 02/15/2009 Page 64 of 72


APPENDIX C: Labels<br />

APPENDIX C: Labels<br />

Rev. 1.0 - 02/15/2009 Page 65 of 72


APPENDIX C: Labels<br />

PLATE ON THE NEWTOM <strong>VG</strong>i<br />

Position: dx upright of the scanning unit, lower side<br />

Rev. 1.0 - 02/15/2009 Page 66 of 72


APPENDIX C: Labels<br />

PLATE ON THE CONTROL BOX<br />

Serial Number :<br />

QR s.r.l. – Via Silvestrini, 20<br />

37135 Verona Italy<br />

+39 045 8202727<br />

Model no: <strong>NewTom</strong> <strong>VG</strong><br />

CONTROL BOX<br />

Manufactured :<br />

Power : 100 / 115 V ~ 10A<br />

200 / 215 / 230 / 240 V ~ 5A<br />

50/60Hz<br />

WARNING: This x-ray unit may be dangerous<br />

to patient and operator unless safe exposure<br />

factors and operating instructions are observed.<br />

WARNING: X-RAYS – ATTENTION: XRAY<br />

ON WHEN EQUIPMENT IN OPERATION<br />

UNAUTHORIZED USE IS STRICTLY<br />

PROHIBITED<br />

Posizione: left side of the control box<br />

CONTROL BOX LABEL<br />

(devices destined at the Chinese market only)<br />

产 品 名 称 口 腔 X 射 线 计 算 机 体 层 摄 影 设 备<br />

产 品 型 号 <strong>NewTom</strong> <strong>VG</strong><br />

系 统 编 号 <strong>VG</strong>XXXXXX<br />

运 行 方 式<br />

间 歇 加 载 连 续 运 行<br />

输 入 电 压<br />

输 入 功 率<br />

220AC+/-10%<br />

50/60Hz<br />

瞬 时 :5A 稳 态 : 0.75A<br />

制 造 商 QR s.r.l. 意 大 利<br />

生 产 地 址 Via Silvestrini,20 - 37135 维 罗 纳 意 大 利<br />

生 产 日 期 YYYY-MM<br />

安 全 标 志<br />

Position: left side of the control box, on the right of the control box plate<br />

Rev. 1.0 - 02/15/2009 Page 67 of 72


APPENDIX C: Labels<br />

LABELS ON THE X-RAY SOURCE<br />

X-Ray source with rotating anode<br />

Inverter<br />

Position: on the structure of the X-Ray source and on the Inverter structure.<br />

Rev. 1.0 - 02/15/2009 Page 68 of 72


APPENDIX C: Labels<br />

DEVICE TRADE NAME LABEL<br />

Position: on the cover of the scanning unit in frontal and centered position<br />

Rev. 1.0 - 02/15/2009 Page 69 of 72


APPENDIX C: Labels<br />

LABEL ON THE BRASS PLATE OF BEAM LIMITER (Full Field)<br />

QR s.r.l. – Italy<br />

P.N. 001092<br />

S.N. ______________<br />

6.73 mm Al @<br />

70<br />

kV<br />

Position: in the center of the brass plate<br />

LABEL ON SMALL FIELD BEAM LIMITER (Zoom)<br />

QR s.r.l. – Italy<br />

P.N. 001067<br />

S.N. ______________<br />

Position: on the lead plate<br />

EQUIVALENT RADIATION LABEL<br />

QR s.r.l. – Italy<br />

P.N. 001065<br />

S.N. ______________<br />

2.34mm @ 70kV<br />

Position: inside the plastic cover of the rotating arm, opposite the x-ray output<br />

LABEL INDICATING THE MAIN SWITCH AND INPUT FUSE<br />

ON-OFF<br />

I<br />

0<br />

FUSE:<br />

Condition Ratings<br />

100 - 115 V~ F1=T10A 250V<br />

200 - 240 V~ F1=T5A 250V<br />

CAUTION: For continued<br />

protection against risk of fire,<br />

replace only with same type<br />

and rating of fuse.<br />

Position: left side of the control box<br />

Rev. 1.0 - 02/15/2009 Page 70 of 72


APPENDIX C: Labels<br />

LABEL WITH CONNECTOR INDICATION<br />

AUX<br />

CAN-BUS<br />

Position: left side of the control box, below<br />

WARNING LABELS FOR DEVICES WITH LASER<br />

CLASS 1 LASER PRODUCT<br />

Wavelenght:630-680 nm Max.Output


APPENDIX C: Labels<br />

HEAD SUPPORT LABELS<br />

Posizion: on the covers of the head support<br />

Rev. 1.0 - 02/15/2009 Page 72 of 72

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!