IEC61223: Acceptance Test<br />

9.7 IEC 61223-3-5 reference table<br />

This table supplies reference values and reference conditions to be used when applying the IEC 61223-3-5<br />

as acceptance test, when specified by contractual agreements.<br />

5.1 Positioning of the<br />

patient support<br />

N.A.<br />

5.2 Patient positioning<br />

accuracy<br />

5.2.1 Axial patient positioning<br />

accuracy<br />

±10mm from isocenter<br />

5.3 Tomographic section<br />

thickness<br />

5.2.2 Sagittal and coronal patient<br />

positioning light accuracy<br />

5.3.1 Tomographic section<br />

thickness for axial scanning<br />

5.3.2 Tomographic section<br />

thickness for helical scanning<br />

±10mm from isocenter<br />

Loading factors: automatically selected by the<br />

software.<br />

Thickness<br />

Raccomanded criteria<br />

5.0 mm ±1.0 mm<br />

1.0 mm ±0.5 mm<br />

0.5 mm -0 +0.5 mm<br />

N.A.<br />

5.4 Dose<br />

Loading factors: automatically selected by the software.<br />

Dose Mode: Standart Dose<br />

CTDI FREE AIR<br />

CTDIvol<br />

Full Mode CTDI 100w<br />

Zoom Mode CTDI 100w<br />

HiRes Zoom Mode CTDI 100w<br />

N.A.<br />

Equal to CTDI 100w<br />

2.4 mGy ±30%<br />

2.0 mGy ±30%<br />

7.3 mGy ±30%<br />

Loading factors: automatically selected by the software.<br />

Dose Mode: Boosted Dose<br />

CTDI FREE AIR<br />

CTDIvol<br />

Full Mode CTDI 100w<br />

Zoom Mode CTDI 100w<br />

N.A.<br />

Equal to CTDI 100w<br />

4.5 mGy ±30%<br />

3.3 mGy ±30%<br />

5.5 Noise, mean CT<br />

number and uniformity<br />

Loading factors: automatically selected by the software.<br />

Values are calculated as Arbitrary Unit (a.u.) because of the machine is not designed to<br />

give HU.<br />

Measured with Catphan 500.<br />

Dose Mode: Standart Dose<br />

Standard deviation (water)<br />

Full Mode 340 a.u. ± 80<br />

Zoom Mode 340 a.u. ± 80<br />

HiRes Zoom Mode 400 a.u. ± 80<br />

Rev. 1.0 - 02/15/2009 Page 41 of 72