Preliminary procedures<br />

5 Preliminary procedures<br />

This chapter describes the procedures to perform compulsorily before starting scanning patients.<br />

These procedures are three:<br />

• X-Ray source conditioning;<br />

• Daily check;<br />

• Blank acquisition;<br />

This chapter describes also some required and/or useful functions:<br />

• Beam limiter test<br />

• Completely lift down the rotating arm<br />

X-Ray source conditioning and Blank acquisition must be performed every two weeks (14 days), while Daily<br />

check is mandatory every day, before patients acquisitions.<br />

If these procedures are not performed the software will disable the scan functions.<br />

Operating modes are described in detail in the same named chapter from the "Software <strong>Manual</strong>".<br />

5.1 X-Ray source conditioning<br />

The conditioning procedure is needed to prepare the x-ray source for its routine working.<br />

The software asks to perform it every two weeks. It is suggested to perform the conditioning also if the<br />

device does not operate for days.<br />

REMOVE<br />

Before starting the procedure, please verify that the<br />

scan area is completely free from objects.<br />

Remove the head support.<br />

At the end of conditioning process, the software will automatically starts Daily check and Blank acquisition<br />

procedures.<br />

5.2 Daily Check<br />

By mean of the Daily Check procedure the system verifies that each every part of itself properly works.<br />

Rev. 1.0 - 02/15/2009 Page 28 of 72