Safety and maintenance of the device<br />

3.5.2 Cleaning<br />

3.5.2.1 Scanner and control box<br />

The device must be cleaned with a wet cloth and a soft solution. The cloth should be wet before applying it<br />

on the scanner surface.<br />

WARNING:<br />

Do not apply cleaning solutions directly on the device surface.<br />

3.5.2.2 <strong>Works</strong>tation and periphera ls<br />

Follow the manufacturer's instructions for cleaning the workstation and its peripheral units. If these are<br />

missing, please refer to the p<strong>rev</strong>ious paragraph.<br />

To clean the display use a wet cloth with alcohol or a solution made up of 1/3 to 2/3 distilled water and<br />

alcohol.<br />

NOTE:<br />

For further information about safety and maintenance of the device please contact<br />

your local distributor.<br />

3.5.3 Disinfection<br />

Despite the device do not require any disinfection, it is suggested to use standard disinfection product for the<br />

cleansing of the head support.<br />

3.5.4 Sterilization<br />

No sterilization is required for the normal use of the device.<br />

3.6 Transport and storage<br />

During transport and storage the following conditions must be observed:<br />

Transport and storage temperature:<br />

Transport and storage humidity:<br />

-20° to +70° (Ce lsius)<br />

min 10%, max 85% not condensing.<br />

Do not expose to acids, saltiness, rain.<br />

Rev. 1.0 - 02/15/2009 Page 18 of 72