Preliminary procedures<br />

From the NNT software main toolbar select Scan <br />

Invalidate Blank.<br />

At the next detector field selection the NNT software<br />

will ask a new blank acquisition<br />

5.4 Beam limiter test<br />

NOTE:<br />

This procedure applies only to devices with multiple FOV.<br />

From the NNT software main toolbar select Tools Scanner Test. The service window will appear.From<br />

the service window toolbar select the Tools Beam Limiter Test. To verify hires field lead collimator<br />

position:<br />

Press “Beam limiter test Acquisition” button.<br />

Verify the acquired image: the light gray rectangle must be vertically center between the two red lines.<br />

5.5 Completely lift down the rotating arm<br />

This function is useful to completely move down the rotating arm. For example during transport , in case of<br />

mobile application , is suggested to position the arm completely down.<br />

WARNING:<br />

Please, verify there are no objects that can obstruct the downward movement of<br />

the arm before using this function!<br />

To completely lift down the rotating arm follow the steps below:<br />

1. From the NNT software main toolbar select Tools Scanner Test. The service window will appear.<br />

2. From the service window toolbar select the Tools Lift down the rotating arm.<br />

Rev. 1.0 - 02/15/2009 Page 30 of 72