Preliminary procedures<br />
From the NNT software main toolbar select Scan <br />
Invalidate Blank.<br />
At the next detector field selection the NNT software<br />
will ask a new blank acquisition<br />
5.4 Beam limiter test<br />
NOTE:<br />
This procedure applies only to devices with multiple FOV.<br />
From the NNT software main toolbar select Tools Scanner Test. The service window will appear.From<br />
the service window toolbar select the Tools Beam Limiter Test. To verify hires field lead collimator<br />
position:<br />
Press “Beam limiter test Acquisition” button.<br />
Verify the acquired image: the light gray rectangle must be vertically center between the two red lines.<br />
5.5 Completely lift down the rotating arm<br />
This function is useful to completely move down the rotating arm. For example during transport , in case of<br />
mobile application , is suggested to position the arm completely down.<br />
WARNING:<br />
Please, verify there are no objects that can obstruct the downward movement of<br />
the arm before using this function!<br />
To completely lift down the rotating arm follow the steps below:<br />
1. From the NNT software main toolbar select Tools Scanner Test. The service window will appear.<br />
2. From the service window toolbar select the Tools Lift down the rotating arm.<br />