Safety and maintenance of the device<br />

Maintenance agreement<br />

The device should be periodically checked: please contact the manufacturer or distributor to discuss for a<br />

maintenance contract.<br />

System testing check list<br />

The following check list specifies the suggested time spaces for the system controls. For further information<br />

please contact Your local distributor.<br />

Responsible Part Activity Frequency<br />

<strong>User</strong> Global system QA phantom check Weekly<br />

Radioprotection<br />

expert (RPE) Global system Radiological routine In charge of the<br />

RPE<br />

Error Log Check 12 months<br />

Every external part Damage check 12 months<br />

Technical staff Emergency button Emergency test 12 months<br />

Electrical functioning Check 12 months<br />

Mechanical functioning Check 12 months<br />

3.5 Cleaning and disinfection<br />

This chapter gives general information about cleaning and disinfection of the device.<br />

WARNING:<br />

Power the device off before any cleaning.<br />

Some cleaning or disinfecting products may damage Your health. Such products concentration in the air<br />

should never exceed the local laws' limits. In this aim always follow the manufacturer instructions and be<br />

sure that the room is well airy.<br />

3.5.1 Chemical agents to avoid<br />

• Do not use spray or solvent which may enter the device, causing damage electrical parts or forming<br />

inflammable mixtures. This could put in danger the patient, the user and the environment.<br />

• Do not use abrasive liquids such as acetone, gas oil or alcohol (exception: alcohol can be used for the<br />

cleaning of the workstation display). These substances can deteriorate surfaces and materials that are<br />

responsible for the device proper working.<br />

• Do not use substances that may release ammonia due to dissociation or decomposition. Ammonia can<br />

corrode.<br />

• Do not use cleaning products containing silicone: it tends to accumulate and possibly damage electrical<br />

contacts.<br />

Rev. 1.0 - 02/15/2009 Page 17 of 72