2.3 Checking the Vacuum<br />

Fit the resin feed system to the resin <strong>and</strong> sample chambers, as shown in the diagram for<br />

section 3.1.3, ensuring that the O‐rings, shown in section 3.1.1, are in place contacting the<br />

top of the glass lid (16). Check visually that the O‐rings on the sample chamber (1) are in<br />

place <strong>and</strong> that the resin chamber (2) is sitting squarely under the glass lid (16) on its spring<br />

loaded base. Ensure that the resin <strong>and</strong> sample chamber valves (10, 12) <strong>and</strong> the resin <strong>and</strong><br />

sample chamber vents (11, 13) are closed.<br />

Switch on the machine using the mains switch at the side by the electrical inlet. The pump<br />

will start <strong>and</strong> the mains indicator light (8) on the pirani gauge (9) will glow green. The pirani<br />

gauge (9) will progressively reduce in reading to the ultimate vacuum achieved by the pump<br />

over a period of up to half a minute. This should be in the region 10 ‐2 to 5x10 ‐2 mbar.<br />

Slowly open the sample chamber valve (12) ‐ the sample chamber (1) will start to pump<br />

down. On a new machine the vacuum will gradually build over a period which may be as<br />

long as half an hour, depending on water vapour etc., which has accumulated in transit<br />

(after this first pump down the sample chamber should evacuate in under five minutes). If<br />

the vacuum fails to start building, ensure that all O‐rings are sitting on clean surfaces.<br />

Repeat this check after opening the resin chamber valve (10), so that both chambers are<br />

evacuated. In both cases the ultimate vacuum reach will depend on atmospheric <strong>and</strong> other<br />

conditions but should be in the region of 5x10 ‐2 mbar to 10 ‐1 mbar.<br />

2.4 Health & Safety<br />

When operating this unit the operator must be aware of the risks involved when working<br />

with resins <strong>and</strong> vacuum pressure.<br />

a) The unit is operated under sufficient vacuum pressure to accelerate the exothermic<br />

reaction in the resin when out gassing. Care must be taken to slowly control the rate<br />

of expansion of the resin bubbling up or effervescing to retain it inside the mixing<br />

vessel. This is done by manually adjusting rate of air removal via the vacuum/air<br />

control valves.<br />

b) The vacuum pump exhaust port emits oil vapour <strong>and</strong> must be vented correctly<br />

through a suitable oil mist filter (see separate vacuum pump manual if supplied with<br />

the machine) or to the facility exhaust system to prevent oil mist in the working<br />

area.<br />

c) The system is not sensitive to vibration <strong>and</strong> can be installed on a st<strong>and</strong>ard bench<br />

which will support the weight (74kg).<br />

<strong>Logitech</strong> <strong>Ltd</strong>, <strong>IU30</strong> <strong>Impregnation</strong> <strong>Unit</strong> – Manual BE‐03‐02‐2 –version v0211 ‐ Page 10