In any of the following circumstances, we recommend that you fit an oil mist filter to the pump outlet:<br />

� If you use the pump with the gas ballast control open (in position ‘I’ or position ‘II’).<br />

A652–01–880 Issue S<br />

� If you operate the pump with an inlet pressure greater than 10 mbar (1 x 10³ Pa) for extended periods.<br />

� If you frequently pump down from atmospheric pressure.<br />

The oil mist filter will trap the oil exhausted from the pump; you can reuse the oil if it is not contaminated.<br />

3.10 Leak-test the system<br />

Leak-test the system <strong>and</strong> seal any leaks found after you have installed the RV pump, to prevent leakage of substances<br />

out of the system <strong>and</strong> leakage of air into the system.<br />

© Edwards Limited 2007. All rights reserved. Page 25<br />

Edwards <strong>and</strong> the Edwards logo are trade marks of Edwards Limited.<br />

Installation