<strong>Operation</strong><br />

A652–01–880 Issue S<br />

4.8 Shut-down<br />

We recommend, as described in the procedure below, that you decontaminate the oil before you shut down the<br />

pump; this will prevent damage to the pump by the contaminates in the oil.<br />

1. Refer to Section 4.6 <strong>and</strong> decontaminate the oil, as required.<br />

2. Close the vacuum system isolation-valve (if not already closed).<br />

3. Close gas-ballast (that is, set the gas-ballast control to position ‘0’).<br />

4. On single-phase pumps, use the on/off switch to switch off the pump.<br />

5. Switch off the electrical supply to the pump.<br />

Page 32 © Edwards Limited 2007. All rights reserved.<br />

Edwards <strong>and</strong> the Edwards logo are trade marks of Edwards Limited.