5.7 Clean the oil-level sight-glass<br />

Refer to Figure 9 for the item numbers in brackets.<br />

1. Drain the oil as described in Section 5.4.<br />

A652–01–880 Issue S<br />

2. Undo the two screws (1) <strong>and</strong> remove the bezel (2), the sight-glass (3) <strong>and</strong> the ‘O’ ring (4) from the oil-box (5).<br />

3. Clean the screws, bezel <strong>and</strong> sight-glass with a suitable cleaning solution.<br />

4. Wipe the ‘O’ ring with a clean, dry, lint-free cloth.<br />

5. Wipe the sight-glass recess in the oil-box with the cloth.<br />

6. Refit the ‘O’ ring, sight-glass <strong>and</strong> bezel <strong>and</strong> secure with the two screws.<br />

7. Refill the pump with oil as described in Section 5.4.<br />

8. Check that the sight-glass does not leak.<br />

Figure 9 - Sight-glass assembly<br />

1. Screws (2 off M6 x 20)<br />

2. Bezel<br />

3. Sight-glass<br />

4. ‘O’ ring<br />

5. Oil-box<br />

© Edwards Limited 2007. All rights reserved. Page 37<br />