7.1 Actions that Occur when the Gas Solenoid Keys are Pressed<br />
Purge and<br />
Gas check<br />
mode<br />
Number of<br />
probes Gas option<br />
Automatic Single Gas 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will start the timed gas<br />
check cycle on solenoid #1 to probe #1<br />
Pressing the GAS 2/ PURGE 2 key will start the timed filter<br />
purge cycle on solenoid #2 to probe #1<br />
Automatic Single Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will start the timed gas<br />
check cycle on solenoid #1 to probe #1<br />
Pressing the GAS 2/ PURGE 2 key will start the timed gas<br />
check cycle on solenoid #2 to probe #1<br />
Automatic Dual Purge 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will start the timed filter<br />
purge cycle on solenoid #1 to probe #1<br />
Pressing the GAS 2/ PURGE 2 key will start the timed filter<br />
purge cycle on solenoid #2 to probe #2<br />
Automatic Dual Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will start the timed gas<br />
check cycle on solenoid #1 to probe #1<br />
Pressing the GAS 2/ PURGE 2 key will start the timed gas<br />
check cycle on solenoid #2 to probe #2<br />
<strong>Manual</strong> Single Purge 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />
purge probe #1 for as long as the key is pressed<br />
Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />
purge probe #1 for as long as the key is pressed<br />
<strong>Manual</strong> Single Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />
pass calibration gas to probe #1 for as long as the key is<br />
pressed<br />
Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />
pass calibration gas to probe #1 for as long as the key is<br />
pressed<br />
<strong>Manual</strong> Dual Purge 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />
purge probe #1 for as long as the key is pressed<br />
Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />
purge probe #2 for as long as the key is pressed<br />
<strong>Manual</strong> Dual Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />
pass calibration gas to probe #1 for as long as the key is<br />
pressed<br />
Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />
pass calibration gas to probe #2 for as long as the key is<br />
pressed<br />
Refer to the person responsible for the commissioning to find out how the transmitter has been configured.<br />