7.1 Actions that Occur when the Gas Solenoid Keys are Pressed<br />

Purge and<br />

Gas check<br />

mode<br />

Number of<br />

probes Gas option<br />

Automatic Single Gas 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will start the timed gas<br />

check cycle on solenoid #1 to probe #1<br />

Pressing the GAS 2/ PURGE 2 key will start the timed filter<br />

purge cycle on solenoid #2 to probe #1<br />

Automatic Single Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will start the timed gas<br />

check cycle on solenoid #1 to probe #1<br />

Pressing the GAS 2/ PURGE 2 key will start the timed gas<br />

check cycle on solenoid #2 to probe #1<br />

Automatic Dual Purge 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will start the timed filter<br />

purge cycle on solenoid #1 to probe #1<br />

Pressing the GAS 2/ PURGE 2 key will start the timed filter<br />

purge cycle on solenoid #2 to probe #2<br />

Automatic Dual Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will start the timed gas<br />

check cycle on solenoid #1 to probe #1<br />

Pressing the GAS 2/ PURGE 2 key will start the timed gas<br />

check cycle on solenoid #2 to probe #2<br />

<strong>Manual</strong> Single Purge 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />

purge probe #1 for as long as the key is pressed<br />

Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />

purge probe #1 for as long as the key is pressed<br />

<strong>Manual</strong> Single Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />

pass calibration gas to probe #1 for as long as the key is<br />

pressed<br />

Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />

pass calibration gas to probe #1 for as long as the key is<br />

pressed<br />

<strong>Manual</strong> Dual Purge 1 & Purge 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />

purge probe #1 for as long as the key is pressed<br />

Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />

purge probe #2 for as long as the key is pressed<br />

<strong>Manual</strong> Dual Gas 1 & Gas 2 Pressing the GAS 1/ PURGE 1 key will turn on solenoid #1 to<br />

pass calibration gas to probe #1 for as long as the key is<br />

pressed<br />

Pressing the GAS 2/ PURGE 2 key will turn on solenoid #2 to<br />

pass calibration gas to probe #2 for as long as the key is<br />

pressed<br />

Refer to the person responsible for the commissioning to find out how the transmitter has been configured.<br />

January 2012<br />

26 <strong>1735</strong> Water Vapour Transmitter