1.2 Heater Warning<br />
The probe heater is supplied with mains voltage. This supply has electrical shock danger to maintenance<br />
personnel. Always isolate the transmitter before working with the probe.<br />
The EARTH wire (green) from the probe must always be connected to earth.<br />
The heater is supplied from the mains power directly, and the temperature is controlled at 720°C (1328°F).<br />
The outside of the process end of the probe can get to temperatures that are dangerous to touch. Wear<br />
insulating gloves when handling a probe that has been on.<br />
January 2012<br />
6 <strong>1735</strong> Water Vapour Transmitter