Chapter 33.4 Configuring the Coulochem III Analyst DriverThe Coulochem III can be configured in either DC or pulsed mode. In DC mode, 1 or 2channels can be utilized, along with an optional guard cell and thermal organizer. Operationin pulsed mode uses a single channel and does not use the guard cell.The driver must be configured to reflect actual hardware as well as experimental preferencesprior to use with the Analyst program for data acquisition.1. From the driver menu, select Instrument > Coulochem III Configuration.Figure 3.4. The Coulochem III Configuration Dialog2. From the configuration dialog, Figure 3.4, select the appropriate options for each ofthe following parameters:• CC III operating modeSelect either DC or Pulse mode.• Comm portSelect USB if connected to the Coulochem III through an available USB port.If connected though a serial port, select either COM1, COM2, COM3 orCOM4, depending on which ports are available. (If unsure, start with COM1.If this causes problems later, return to the configuration and try COM2, COM316 Coulochem ® III Analyst Driver – Version 1.0User’s Manual