Updating the Coulochem III FirmwareWhen properly connected through the USB cable, the message “Connected” appearsnear the top of the utility window, and the utility shows both the serial number and thecurrent firmware (software) version number.4. Click the Update button. When the Open dialog appears, navigate to the ProgramFiles folder (normally on Drive C), then to the ESA Biosciences subfolder and finallyto the Corona-Coulochem Firmware Update Utility folder (i.e., to the folderC:\Program Files\ESA Biosciences\Corona-Coulochem Firmware Update Utility).5. Select the file coulochem #.##.run, where #.## is the version number of the newfirmware update (for example, “coulochem 3.07.run”). If more than one such file isfound, select the one with the highest (latest) version number. Then click Open.6. The update utility will replace all of the existing firmware with the version on theupdate CD. The process should take no more than a few minutes. When finished, themessage “Download complete Coulochem III rebooting” will appear near the top ofthe window and the Coulochem III will reboot automatically.7. Click Exit to close the update utility.Coulochem ® III Analyst Driver – Version 1.0 59User’s Manual