Creating and Modifying Acquisition MethodsE1 – E4 Potentials. Values from -2000 to +2000 mV are allowed. Defaultvalues for potentials E1 thru E4 are:E1: 200 mVE2: -1000 mVE3: 600 mVE4: -100 mVE1 – E4 Pulse Widths. The pulse width for E1 must be at least 5 msec greaterthan the acquisition delay. Maximum value for all pulse widths is 1000 msec.Minimum and default values are:E1: AD+5 msec minimum 500 msec defaultE2: 4 msec minimum 10 msec defaultE3: 0 msec minimum 1 msec defaultE4: 0 msec minimum 10 msec defaultFull scale gain range. The full scale gain range runs from 10 pC to 10 mC insteps of 1, 2, 5, 10, 20, 50, 100, 200 and 500. Set this value above the highestsignal expected over the course of the experiment.Signal output voltage. Select from values of -1 V, -100 mV, +100 mVand +1 V. Default is +1 V.% Baseline offset. Offset values from -50% to +50% are allowed. Defaultis 0 %.Filter setting. Select None, Low, Med or High. Use the lowest filter neededto reduce output noise. Default is None.Coulochem ® III Analyst Driver – Version 1.0 29User’s Manual