5. PARAMETER SETUP[F20 PRESS UNIT] Pressure unitSelect pressure unit from MPa abs(0), kPa abs(1), bar abs(2),kg/cm 2 a(3), psia(4).[F21 PRESS f] Pressure value fSet absolute pressure values at operating condition.[F22 PRESS b] Pressure value bSet absolute pressure values at normal/standard Condition.[F23 DAVIATION] Daviation factorSet the daviation factor.[F24 STD/NOR UNIT] Standard/Normal unitSelect Volumetric unit at standard/normal condition FromNm3(0), k Nm3(1), M Nm3(2), Nl(3), Sm3(4)k Sm3(5), M Sm3(6), Sl(7), scf(8), k scf(9), M scf(10)N: NormalS: StandardThe following item should be done in case ofwhich F10 is LIQUID:MASS[F26 DENSITY UNIT] Density unitSelect density unit from kg/m3(0), lb/cf(1), lb/USgal(2),lb/UKgal(3).[F27 DENSlTY b] Density bSet density value at standard condition.[F28 TEMP UNIT] Temperature unitSelect temperature unit from deg C(0), deg F(1).[F29 TEMP b] Temperature bSet temperature value at standard condition[F50 TEMP DAMPING] Temperature dampingSet temperature damping, 0 to 99sec.[F52 CABLE LENGTH] Cable length of signalcable(DYC)Set cable length(m) of signal cable.In case of the integral type, cable length is set in 0m.<strong>IM</strong>PORTANTBe sure to set this parameter to correct temperaturemeasurement error, occured by cable length.[F55 A/OUT SELECT] Analog out selectSelect the analog output from FLOW(0), TEMP(1).The following item should be done in case ofwhich F55 is TEMP[F56 TEMP 0%] Temperature at 0%Set temperature value at 0%.[F57 TEMP 100%] Temperature at 100%Set temperature value at 100%.[F58 TEMP ERR OUT] Output selection ofthermometer errorSelect output function when thermometer error from 0%(0),OUTL<strong>IM</strong>IT(H)(1), TEMP f.In case of OUT L<strong>IM</strong>IT(H), it is based on parameter “D30”[F30 1st coef] 1st coefficientSet lst temperature coefficient using the density correction.[F31 2nd coef] 2nd coefficientSet lst temperature coefficient using the density correction.[F32 MASS UNIT] Mass unitSelect mass flow rate unit from kg(0), t(1), lb(2), k lb(3).[F35 T<strong>IM</strong>E UNIT] Time unitSelect time unit from /s(0), /m(1), /h(2), /d(3).[F40 FLOW SPAN] Flow spanSet span flow rate, 0 to 32000.[F45 DAMPING] Flow dampingSet flow damping, 0 to 99sec.5-14 <strong>IM</strong> <strong>01F06A00</strong>-<strong>01EN</strong>

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!