❏ Inaccurate Readings✓ Verify spindle, speed and model selection.✓ If % readings are under-range (less than 10%), the display will flash; change spindle and/orspeed.✓ “EEE” on the digital display means the unit is over-range (greater than 100%); reduce speedand/or change spindle.✓ Verify test parameters: temperature, container, volume, method. Refer to:• “More Solutions to Sticky Problems”, Section III• “<strong>DV</strong>-<strong>II+</strong> Programmable Viscometer <strong>Operating</strong> Instructions, Appendix C, “Variables inViscosity Measurements.✓ Perform a calibration check; follow the instructions in Appendix E.✓ Verify tolerances are calculated correctly.✓ Verify the calibration check procedures were followed exactly.If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix K fordetails on “How to Return Your Viscometer”.❏ Viscometer Will Not Return to Zero✓ Viscometer is not level• Check with spindle out of the sample• Adjust the laboratory stand✓ Pivot point or jewel bearing faulty• Perform calibration check• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair❏ Display Reading Will Not Stabilize✓ Special characteristic of sample fluid. There is no problem with the viscometer.• Refer to Appendix C✓ Check for erratic spindle rotation.• Verify power supply• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair✓ Bent spindle or spindle coupling.• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair✓ Temperature fluctuation in sample fluid• Use temperature bath control.Brookfield Engineering Labs., Inc. Page 69 Manual No. M/97-164-F1 102