TABLE 21-3. THE EFFECT OF INCREASING COLLIMATOR-TO-OBJECT DISTANCEON COLLIMATOR PERFORMANCE PARAMETERSCollimator Spatial resolution" Efficiency Field size MagnificationParallel hole Decreases Approximately Constant Constant (m = 1.0)constantConverging Decreases Increases Decreases Increases (m >1 atcollimator surface)Diverging Decreases Decreases Increases Decreases (m