13.07.2015 Views

Bush__The_Essential_Physics_for_Medical_Imaging - Biomedical ...

Bush__The_Essential_Physics_for_Medical_Imaging - Biomedical ...

Bush__The_Essential_Physics_for_Medical_Imaging - Biomedical ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Scintillation camera planarSPECTPETDigital fluoroscopy, cardiac catheter labComputed radiography, digitizedchest filmsMammography (18 x 24 cm) or (24 x 30 cm)X-ray CTMRIUltrasound64 2 or 128 264 2 or 128 2128 2512 2 or 1024 2Typically 2,000 x 2,500Typically 1,800 x 2,300 to4,800 x 6,000512 264 2 to 1,024 2512 28 or 168 or 16168 to 1210-12CT, computed tomography; MRI, magnetic resonance imaging; PET,positron emission tomography; SPECT,single photon emission computed tomography.a digital image in four different pixel <strong>for</strong>mats. <strong>The</strong> number in each pixel is convertedinto a visible light intensity when the image is displayed on a video monitor. Typicalimage <strong>for</strong>mats used in radiology are listed in Table 4-7.<strong>Imaging</strong> modalities with higher spatial resolution require more pixels per imageso the image <strong>for</strong>mat does not degrade the resolution. In general, an image <strong>for</strong>matshould be selected so that the pixel size is on the order of the size of the smallest objectto be seen. In the case of a fluoroscope with a 23 cm field of view, the 512 2 pixel <strong>for</strong>matwould be adequate <strong>for</strong> detecting objects as small as about a half a millimeter insize. To detect objects half this size, a larger <strong>for</strong>mat, such as 1024 by 1024, should beselected. When it is necessary to depict the shape of an object, such as a microcalcificationin x-ray mammography, an image <strong>for</strong>mat much larger than that needed tomerely detect the object is required. Figure 4-9 shows the degradation of spatial resolutioncaused by using too small an image <strong>for</strong>mat. <strong>The</strong> penalty <strong>for</strong> using larger <strong>for</strong>matsis increased storage and processing requirements and slower transfer of images.<strong>The</strong> largest number that can be stored in a single pixel is determined by thenumber of bits or bytes used <strong>for</strong> each pixel. If 1 byte (8 bits) is used, the maximalnumber that can be stored in one pixel is 255 (2 8 - 1). If2 bytes (16 bits) are used,the maximal number that can be stored is 65,535 (2 16 - 1). <strong>The</strong> amount of contrastresolution provided by an imaging modality determines the number of bits requiredper pixel. <strong>The</strong>re<strong>for</strong>e, imaging modalities with higher contrast resolution requiremore bits per pixel. For example, the limited contrast resolution of ultrasound usuallyrequires only 6 or 7 bits, and so 8 bits are commonly used <strong>for</strong> each pixel. Onthe other hand, x-ray CT provides high contrast resolution and 12 bits are requiredto represent the full range of CT numbers. Figure 4-10 shows the degradation ofspatial resolution caused by using too few bits per pixel.Pixel size is determined by dividing the distance between two points in thesubject being imaged by the number of pixels between these two points in theimage. It is approximately equal to the field of view of the imaging device dividedby the number of pixels across the image. For example, if a fluoroscope has a 23-cm (9-inch) field of view and the images are acquired in a 512-by- 512 <strong>for</strong>mat,then the approximate size of a pixel is 23 cm/512 = 0.45 mm, <strong>for</strong> objects at theface of the image intensifier.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!