TABLE 5-4. ROOT-MEAN-SQUAREVOLTAGE (Vrms) AS A FRACTION OFPEAK VOLTAGE {V peak)V,msas a fractionVpeakofSingle-phaseThree-phase six-pulseThree-phase 12-pulseHigh-frequencyConstant-potential0.710.950.990.95-0.991.00generator wave<strong>for</strong>m and cable capacitance properties, is used in the equation todetermine the power delivered over the lOO-msec period. For all but constantpotentialgenerators, the average kV is less than the kVp (due to voltage ripple), andit is substantially less <strong>for</strong> a single-phase generator. Exact determination of the powerrating requires multiplication by the "root-mean-square" voltage (V,ms), as listed inTable 5-4 <strong>for</strong> various x-ray generator voltage wave<strong>for</strong>ms. (V,msis the constant voltagethat would deliver the same power as the varying voltage wave<strong>for</strong>m). This correctionis usually ignored <strong>for</strong> all but the single-phase generator.<strong>The</strong>se two examples illustrate the power delivery capability of the generator andpower deposition tolerance of the focal spot. Table 5-5 classifies the x-ray generator interms of power output, and Table 5-6 lists typical x-ray tube focal spot power ratings.X-ray generators have circuits to prohibit combinations of kVp, mA, and exposuretime that will exceed the single-exposure power deposition tolerance of the focal spot.Typical PowerPower Level RequirementsGenerator Type Usage Applications (kW)Single-pulse, one-phase Very low Dental and handheld :Qfluoroscopy unitsTwo-pulse, one-phase Low to medium General fluoroscopy and ~10 and ~50radiographySix-pulse, three-phase Medium to high Light special procedures ~30 and ~100Twelve-pulse, three-phase High Interventional and cardiac ~50 and ~150angiographyConsta nt-potentia I High Interventional and cardiac ~80 and ~200angiographyHigh-frequency All ranges All radiographic procedures ~2 and ~150

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!