Directions for Use Chapter 6 Reference 59<br />
Eight-bay battery charger specifications<br />
Characteristic Specification<br />
Reorder Number<br />
Universal Battery Charger<br />
Active C<br />
Functional Specifications<br />
008-0651-XX<br />
Capacity<br />
Protection Classifications<br />
Eight charging bays; able to charge four (Lithium Ion) batteries<br />
simultaneously.<br />
a<br />
Duty Cycle Continuous<br />
Type of Protection Against Electric<br />
Shock<br />
Class I, (Protectively Earthed) with Double Insulation<br />
Degree of Protection Against Harmful<br />
Ingress of Water<br />
For ordinary, indoor locations only.<br />
Method of Disinfection Not suitable for autoclaving. (See cleaning instructions on “Inspect <strong>and</strong><br />
clean the monitor <strong>and</strong> accessories” on page 45.)<br />
Flammable Anesthetics<br />
Environmental Specifications<br />
Not suitable for use with flammable anesthetics.<br />
Operating Temperature 0° to 40° C<br />
Shipping <strong>and</strong> Storage Temperature -20° to 60° C<br />
Operating Altitude -2,000 to 15,000 feet (-610 to 4,572 m)<br />
Shipping <strong>and</strong> Storage Altitude -2,000 to 40,000 feet (-610 to 12,192 m)<br />
Operating Relative Humidity 15% to 95%, noncondensing<br />
Shipping, Storage Relative Humidity 15% to 95%, noncondensing<br />
Shock 30 g<br />
Vibration 0.01g 2 /Hz from 5 to 500 Hz, 30 minutes per axis<br />
Electromagnetic Compatibility (EMC)<br />
Physical Specifications<br />
EN60601-1-2: 2001<br />
Length 15.0 in (38.1 cm)<br />
Width 9.0 in (22.9 cm)<br />
Height 3.6 in (9.1 cm) including feet<br />
Weight<br />
Electrical Specifications<br />
3.5 lb (1.6 kg)<br />
Rated Input 100 V-240 V AC 600 mA, 50/60 ± 3 Hz, Electrical Class I<br />
Rated Fuses T1.25 A/250V, Time-Delay 5x20mm<br />
Rated Output per charging bay<br />
(Continuous)<br />
8.4 V ± 100 mV dc @1 A max.<br />
Charge Time 4 hours typical for fully discharged battery. Automatic charge termination<br />
when charge is completed, or fault detected.<br />
Output Over-Current Electronic overload protection<br />
Additional Features<br />
LED Indicators<br />
Detachable power cord<br />
LED OFF<br />
Flashing GREEN LED<br />
No battery detected.<br />
1 sec ON, 3 sec OFF<br />
Battery detected, waiting to be charged<br />
1 sec ON, 1 sec OFF<br />
Battery is charged<br />
Continuously ON GREEN LED Battery is charged.<br />
Continuously ON YELLOW LED Battery or charging bay fault.<br />
a. Per EN 60601-1 unless otherwise stated.