Operator’s Manual Addendum<br />
Flow pattern Range:<br />
Flow pattern not selectable when mandatory type is PC or VC+.<br />
New patient value:<br />
Neonatal: Descending ramp<br />
Pediatric: Square<br />
Adult: Square<br />
Flow sensitivity<br />
(V SENS)<br />
High spontaneous<br />
inspiratory time limit<br />
(2TI SPONT)<br />
(Available when Vent<br />
Type is NIV, only)<br />
Table 3. Ventilator settings updates (continued)<br />
Setting Function/Details<br />
Range:<br />
Neonatal: 0.1 L/min to 10 L/min<br />
Pediatric/Adult: 0.2 L/min to 20 L/min<br />
New patient value:<br />
Neonatal: 1.0 L/min<br />
Pediatric: 2.0 L/min<br />
Adult: 3.0 L/min<br />
Range:<br />
Neonatal: 0.4 sec to (1 + (0.1 x IBW)) sec<br />
Pediatric/Adult: 0.4 sec to (1.99 + (0.02 x IBW)) sec<br />
New patient value:<br />
Neonatal: (1 + (0.1 x IBW)) sec<br />
Pediatric/Adult: (1.99 + (0.02 x IBW)) sec<br />
Humidifier volume Function: The empty volume of the currently-installed humidifier.<br />
Range: 100 mL to 1000 mL<br />
Default: 480 mL<br />
Resolution: 10 mL<br />
12 10070808 Rev. C Software Enhancements