Operator’s Manual Addendum<br />

Flow pattern Range:<br />

Flow pattern not selectable when mandatory type is PC or VC+.<br />

New patient value:<br />

Neonatal: Descending ramp<br />

Pediatric: Square<br />

Adult: Square<br />

Flow sensitivity<br />

(V SENS)<br />

High spontaneous<br />

inspiratory time limit<br />

(2TI SPONT)<br />

(Available when Vent<br />

Type is NIV, only)<br />

Table 3. Ventilator settings updates (continued)<br />

Setting Function/Details<br />

Range:<br />

Neonatal: 0.1 L/min to 10 L/min<br />

Pediatric/Adult: 0.2 L/min to 20 L/min<br />

New patient value:<br />

Neonatal: 1.0 L/min<br />

Pediatric: 2.0 L/min<br />

Adult: 3.0 L/min<br />

Range:<br />

Neonatal: 0.4 sec to (1 + (0.1 x IBW)) sec<br />

Pediatric/Adult: 0.4 sec to (1.99 + (0.02 x IBW)) sec<br />

New patient value:<br />

Neonatal: (1 + (0.1 x IBW)) sec<br />

Pediatric/Adult: (1.99 + (0.02 x IBW)) sec<br />

Humidifier volume Function: The empty volume of the currently-installed humidifier.<br />

Range: 100 mL to 1000 mL<br />

Default: 480 mL<br />

Resolution: 10 mL<br />