Operator’s Manual Addendum<br />
Trigger type: Pressure Trigger type: Flow<br />
(Flow triggering is only allowable trigger<br />
type in NIV)<br />
Alarm settings: 4P PEAK (if applicable),<br />
4V ETOT, 4V TE MAND, 4V TE SPONT, INSPIRATION<br />
TOO LONG (alarm limit based on IBW)<br />
D SENS<br />
Changing patient from NIV to INVASIVE Vent Type<br />
Alarm settings: 4PPEAK , 4VE TOT , 4VTE MAND ,<br />
4VTE SPONT default to NIV new patient values<br />
(see Table 4). INSPIRATION TOO LONG alarm<br />
not available.<br />
DSENS setting defaults to OFF.<br />
Table 9 shows automatic settings changes that occur when changing the same<br />
patient from NIV to INVASIVE Vent Type.<br />
Table 9. Automatic settings changes – NIV to INVASIVE<br />
Current NIV setting New INVASIVE setting<br />
Ventilator settings: 2TI SPONT<br />
Alarm settings: 4PPEAK , 4VE TOT , 4VTE MAND ,<br />
4VTE SPONT<br />
D SENS<br />
Table 8. Automatic settings changes – INVASIVE to NIV (continued)<br />
Current INVASIVE setting New NIV setting<br />
Warning<br />
When changing the Vent Type on the same patient, review the automatic<br />
settings changes described in Tables 8 and 9 and adjust appropriately.<br />
36 10070808 Rev. C Software Enhancements<br />
N/A<br />
Alarm settings: Default to new patient<br />
values dependent upon selected INVASIVE<br />
ventilator settings. INSPIRATION TOO LONG<br />
alarm becomes available.<br />
DSENS setting defaults to INVASIVE new<br />
patient value.