Operator’s Manual Addendum<br />

Trigger type: Pressure Trigger type: Flow<br />

(Flow triggering is only allowable trigger<br />

type in NIV)<br />

Alarm settings: 4P PEAK (if applicable),<br />

4V ETOT, 4V TE MAND, 4V TE SPONT, INSPIRATION<br />

TOO LONG (alarm limit based on IBW)<br />

D SENS<br />

Changing patient from NIV to INVASIVE Vent Type<br />

Alarm settings: 4PPEAK , 4VE TOT , 4VTE MAND ,<br />

4VTE SPONT default to NIV new patient values<br />

(see Table 4). INSPIRATION TOO LONG alarm<br />

not available.<br />

DSENS setting defaults to OFF.<br />

Table 9 shows automatic settings changes that occur when changing the same<br />

patient from NIV to INVASIVE Vent Type.<br />

Table 9. Automatic settings changes – NIV to INVASIVE<br />

Current NIV setting New INVASIVE setting<br />

Ventilator settings: 2TI SPONT<br />

Alarm settings: 4PPEAK , 4VE TOT , 4VTE MAND ,<br />

4VTE SPONT<br />

D SENS<br />

Table 8. Automatic settings changes – INVASIVE to NIV (continued)<br />

Current INVASIVE setting New NIV setting<br />

Warning<br />

When changing the Vent Type on the same patient, review the automatic<br />

settings changes described in Tables 8 and 9 and adjust appropriately.<br />

36 10070808 Rev. C Software Enhancements<br />

N/A<br />

Alarm settings: Default to new patient<br />

values dependent upon selected INVASIVE<br />

ventilator settings. INSPIRATION TOO LONG<br />

alarm becomes available.<br />

DSENS setting defaults to INVASIVE new<br />

patient value.