IFU for 840softwareenhancementsaddendumenglish - Covidien
IFU for 840softwareenhancementsaddendumenglish - Covidien
IFU for 840softwareenhancementsaddendumenglish - Covidien
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Operator’s Manual Addendum<br />
Table 10 lists the SNDF message components and their descriptions.<br />
NOTE:<br />
Non-applicable fields will either contain zero or be blank.<br />
Table 10. SNDF response<br />
Component Description<br />
MISCF Response to SNDF command (5 characters)<br />
1253 Number of bytes between and (4 characters)<br />
173 Number of fields between and (3 characters)<br />
Start of transmission character (02 hex)<br />
Field 5 Ventilator time (HH:MM_) (6 characters)<br />
Field 6 Ventilator ID to allow external hosts to uniquely identify each 840<br />
ventilator (18 characters)<br />
Field 7 Date (MMM_DD_YYYY_) (12 characters)<br />
Field 8 Vent Type (NIV______ or INVASIVE_) (9 characters)<br />
Field 9 Mode (A/C___, SIMV__, SPONT_ or BILEVL) (6 characters)<br />
Field 10 Mandatory Type (PC____, VC____, VC+___)<br />
(6 characters)<br />
Field 11 Spontaneous Type (NONE__, PS____, TC____, VS____, PA____)<br />
(6 characters)<br />
Field 12 Trigger Type setting (V-Trig or P-Trig) (6 characters)<br />
Field 13 Respiratory rate setting in bpm (6 characters)<br />
Field 14 Tidal volume setting in L (6 characters)<br />
Field 15 Peak flow setting in L/min (6 characters)<br />
Field 16 O2 % setting (6 characters)<br />
Field 17 Pressure sensitivity setting in cmH2O (6 characters)<br />
Field 18 PEEP/CPAP in cmH2O (6 characters)<br />
Field 19 Plateau setting in seconds (6 characters)<br />
Field 20 Apnea interval setting in seconds (6 characters)<br />
Field 21 Apnea tidal volume setting in L (6 characters)<br />
38 10070808 Rev. C Software Enhancements