01.11.2012 Views

IFU for 840softwareenhancementsaddendumenglish - Covidien

IFU for 840softwareenhancementsaddendumenglish - Covidien

IFU for 840softwareenhancementsaddendumenglish - Covidien

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Operator’s Manual Addendum<br />

Base<br />

message<br />

Table 6. Alarm summary updates (continued)<br />

Urgency<br />

Analysis<br />

message<br />

1P PEAK Low Last breath � set<br />

limit.<br />

Medium Last 3 breaths<br />

� set limit.<br />

High Last 4 or more<br />

breaths � set limit.<br />

1VTE Low Last 2 breaths<br />

� set limit.<br />

Medium Last 4 breaths<br />

� set limit.<br />

High Last 10 or more<br />

breaths<br />

� set limit.<br />

1VE TOT Low VE TOT � set limit.t<br />

<strong>for</strong> � 30s.<br />

Medium VE TOT � set limit<br />

<strong>for</strong> > 30s.<br />

High VE TOT � set limit<br />

<strong>for</strong> > 120s.<br />

1f TOT Low f TOT � set limit <strong>for</strong><br />

� 30s.<br />

Medium fTOT � set limit <strong>for</strong><br />

> 30s.<br />

High fTOT � set limit <strong>for</strong><br />

> 120s.<br />

Remedy<br />

message<br />

Check patient<br />

circuit & ET<br />

tube.<br />

Measured airway<br />

pressure � set limit.<br />

Ventilator truncates<br />

current breath unless<br />

already in exhalation.<br />

Possible dependent<br />

alarms:<br />

3VTE MAND<br />

3V E TOT<br />

1f TOT<br />

Comments<br />

Unchanged. Unchanged.<br />

Unchanged. Unchanged.<br />

Unchanged. Unchanged.<br />

24 10070808 Rev. C Software Enhancements