IFU for 840softwareenhancementsaddendumenglish - Covidien
IFU for 840softwareenhancementsaddendumenglish - Covidien
IFU for 840softwareenhancementsaddendumenglish - Covidien
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Operator’s Manual Addendum<br />
1<br />
2<br />
3<br />
4<br />
1. Vent Type Button: New button used to select between INVASIVE or<br />
NIV.<br />
2. Breath Mode: Only A/C, SIMV, and SPONT modes are allowed with<br />
NIV.<br />
3. Mandatory Type: Only VC and PC are available with NIV.<br />
4. Spontaneous Type: Only PS or NONE are available with NIV when<br />
SIMV or SPONT breath mode is selected.<br />
5. Trigger Type: Only Flow Triggering is available with NIV.<br />
Figure 6. New patient set up screen – NIV<br />
28 10070808 Rev. C Software Enhancements<br />
5