Operator’s Manual Addendum<br />

Table 10. SNDF response (continued)<br />

Component Description<br />

Field 44 Base flow setting in L/min (6 characters)<br />

Field 45 Flow sensitivity setting in L/min (6 characters)<br />

Field 46 PCV inspiratory pressure (PI ) setting in cmH2O (6 characters)<br />

Field 47 PCV inspiratory time (TI) setting in seconds (6 characters)<br />

Field 48 Inspiratory component of I:E ratio setting or High component of H:L<br />

ratio setting (6 characters)<br />

Field 49 Expiratory component of I:E ratio setting or Low component of H:L ratio<br />

setting (6 characters)<br />

Field 50 Constant during rate change setting (I-time, I/E, or E-time) (6 characters)<br />

Field 51 Tube I.D. setting in mm (6 characters)<br />

Field 52 Tube type setting (ET or TRACH) (6 characters)<br />

Field 53 Humidification type setting (Non-Heated Exp, Heated Exp, or HME)<br />

(18 characters)<br />

Field 54 Humidifier volume setting in L (6 characters)<br />

Field 55 O2 sensor setting (Enabled or Disabled) (9 characters)<br />

Field 56 Disconnect sensitivity setting in % or OFF (6 characters)<br />

Field 57 Rise time % setting (6 characters)<br />

Field 58 PAV*+ percent support setting (6 characters)<br />

Field 59 Expiratory sensitivity (ESENS ) setting in % or L/min <strong>for</strong> PA breath type<br />

(6 characters)<br />

Field 60 IBW setting in kg (6 characters)<br />

Field 61 Target support volume (VT SUPP) setting in L (6 characters)<br />

Field 62 High PEEP (PEEPH ) setting in cmH2O (6 characters)<br />

Field 63 Low PEEP (PEEPL ) setting in cmH2O (6 characters)<br />

Field 64 High PEEP time (TH) setting in seconds (6 characters)<br />

Field 65 High spontaneous inspiratory time limit (2TI SPONT ) setting in seconds<br />

(6 characters)<br />

Field 66 Circuit type setting (ADULT, PEDIATRIC, or NEONATAL) (9 characters)<br />

Field 67 Low PEEP time (TL ) setting in seconds (6 characters)<br />

40 10070808 Rev. C Software Enhancements