Operator’s Manual Addendum<br />
• Alarm Silence in Progress and 100% O2 /CAL in Progress indicators (when<br />
active) and the CANCEL buttons <strong>for</strong> them are displayed on the lower GUI<br />
screen.<br />
• The main settings (buttons displayed at the top of the lower screen) can be set<br />
individually or in a batch to allow quick setup.<br />
• Drop-down menus of available selections have been added <strong>for</strong> mode,<br />
mandatory type, spontaneous type, trigger type, apnea mandatory type, and<br />
graphics plot setup.<br />
• The lower screen displays monitored settings if you select or change the settings<br />
that affect them:<br />
� Selecting or changing the volume setting displays the current volume per<br />
weight ratio (VT/IBW or VT SUPP/IBW).<br />
� Selecting or changing the respiratory rate or volume setting displays the set<br />
minute volume (VE SET).<br />
• Expiratory sensitivity (ESENS) is now a primary setting that appears at the top of<br />
the lower screen. ESENS can be set as high as 80% <strong>for</strong> leak management.<br />
• The alarm setting bars on the Alarm Setup screen display the recent range of<br />
the corresponding patient data.<br />
• Circuit disconnect and patient data alarms no longer reset an active alarm<br />
silence.<br />
• Oxygen sensor calibration (100% O2 CAL) can be canceled.<br />
• An automatic expiratory or inspiratory pause maneuver can be canceled by<br />
touching the CANCEL button on the lower GUI display.<br />
• If the ventilator enters idle mode or an occlusion status cycling (OSC) state<br />
when NeoMode is active, the ventilator delivers 40% O2 if available.<br />
• Peak Circuit Pressure (PPEAK) shows the peak inspiratory pressure, and is<br />
updated at the end of each inspiration. Previously, this number represented the<br />
peak pressure <strong>for</strong> the whole breath.<br />
• Mean Circuit Pressure (PMEAN) indicates the average circuit pressure over all<br />
breaths <strong>for</strong> the previous 1-minute interval.<br />
• The O2 sensor alarm is no longer a DEVICE ALERT alarm.<br />
• Various alarm parameters have been changed to reduce the occurrence of<br />
nuisance alarms.<br />
2 10070808 Rev. C Software Enhancements