Operator’s Manual Addendum<br />
Low exhaled<br />
spontaneous tidal<br />
volume<br />
(4V TE SPONT )<br />
Low circuit pressure<br />
(4P PEAK)<br />
Table 4. Alarm settings updates (continued)<br />
Alarm Range<br />
Range:<br />
OFF or<br />
��1 mL<br />
< high exhaled tidal volume limit<br />
and<br />
Neonatal: ��300 mL<br />
Pediatric: ��1000 mL<br />
Adult: ��2500 mL<br />
New patient value (INVASIVE Vent Type): Based on IBW<br />
New patient value (NIV Vent Type): OFF<br />
Resolution:<br />
1 mL <strong>for</strong> 1 mL to 100 mL<br />
5 mL <strong>for</strong> 100 mL to 400 mL<br />
10 mL <strong>for</strong> 400 mL to 2500 mL<br />
Available only during NIV or when VC+ is selected as Mandatory Type<br />
during INVASIVE ventilation.<br />
Range:<br />
NIV: OFF to 2PPEAK -1 cmH2O VC+: PEEP to 2PPEAK -1 cmH2O NOTE:<br />
When VC+ is selected, 4P PEAK can be set to OFF if PEEP is set<br />
to 0.<br />
New patient value: PEEP + 6 cmH2O Resolution:<br />
0.5 cmH2O <strong>for</strong> pressures < 20 cmH2O 1 cmH2O <strong>for</strong> pressures � 20 cmH2O 18 10070808 Rev. C Software Enhancements