Operator’s Manual Addendum<br />

Table 7 summarizes the differences in upper and lower GUI screens between NIV<br />

and INVASIVE ventilation types.<br />

Table 7. Visual differences between NIV and INVASIVE Vent Types<br />

Display NIV INVASIVE Comments<br />

Upper GUI:<br />

Vital patient<br />

data area (top of<br />

upper GUI<br />

screen)<br />

Upper GUI:<br />

More Patient<br />

Data subscreen<br />

Upper GUI:<br />

Alarm area<br />

Lower GUI:<br />

Upper left of<br />

settings area<br />

PPEAK, PMEAN, VTI, I:E, fTOT , VTE ,<br />

VE TOT<br />

Displays current PEEP<br />

value in addition to<br />

same data displayed<br />

in INVASIVE Vent<br />

Type.<br />

NIV and 1T I SPONT (if<br />

set limit is reached)<br />

displayed in yellow<br />

letters.<br />

N A/C<br />

N SIMV<br />

N SPONT<br />

displayed as black<br />

letters in yellow<br />

highlight<br />

PPEAK, PMEAN, PEEP, I:E, fTOT , VTE ,<br />

VE TOT<br />

Data displayed<br />

according to selected<br />

breath mode.<br />

When Vent Type is<br />

NIV, PEEP is displayed<br />

on More Patient Data<br />

subscreen.<br />

Patient data varies<br />

according to selected<br />

breath mode.<br />

N/A If 2 alarms are<br />

simultaneously active,<br />

“NIV” and “1T I SPONT”<br />

are hidden from view<br />

until one alarm is no<br />

longer active.<br />

Breath mode<br />

displayed in white<br />

letters.<br />

Indicator <strong>for</strong> NIV<br />

always present on<br />

lower GUI screen<br />

regardless of alarm<br />

conditions.<br />