Operator’s Manual Addendum<br />
Table 4. Alarm settings updates<br />
Alarm Range<br />
High circuit pressure<br />
(2PPEAK) High exhaled minute<br />
volume<br />
(2VE TOT)<br />
High exhaled tidal<br />
volume limit<br />
(2VTE) High respiratory rate<br />
(2f TOT )<br />
New patient value:<br />
Neonatal: 30 cmH 2O<br />
Pediatric/Adult: 40 cmH 2 O<br />
Range:<br />
OFF or<br />
��0.10 L/min or > low exhaled minute volume limit<br />
and<br />
Neonatal: ��10 L/min<br />
Pediatric: ��30 L/min<br />
Adult: ��100 L/min<br />
New patient value: Based on IBW<br />
Range:<br />
OFF or<br />
> low exhaled spontaneous tidal volume limit<br />
> low exhaled mandatory tidal volume limit<br />
and<br />
Neonatal: 5 mL to 500 mL<br />
Pediatric: 25 mL to 1500 mL<br />
Adult: 25 mL to 3000 mL<br />
New patient value: Based on IBW<br />
Resolution:<br />
1 mL <strong>for</strong> 5 mL to 100 mL<br />
5 mL <strong>for</strong> 100 mL to 400 mL<br />
10 mL <strong>for</strong> 400 mL to 3000 mL<br />
Range:<br />
OFF or<br />
Neonatal: 10 /min to 170 /min<br />
Pediatric/Adult: 10 /min to 110 /min<br />
New patient value: OFF<br />
Resolution: 1 /min<br />
16 10070808 Rev. C Software Enhancements