Operator’s Manual Addendum<br />

Table 4. Alarm settings updates<br />

Alarm Range<br />

High circuit pressure<br />

(2PPEAK) High exhaled minute<br />

volume<br />

(2VE TOT)<br />

High exhaled tidal<br />

volume limit<br />

(2VTE) High respiratory rate<br />

(2f TOT )<br />

New patient value:<br />

Neonatal: 30 cmH 2O<br />

Pediatric/Adult: 40 cmH 2 O<br />

Range:<br />

OFF or<br />

��0.10 L/min or > low exhaled minute volume limit<br />

and<br />

Neonatal: ��10 L/min<br />

Pediatric: ��30 L/min<br />

Adult: ��100 L/min<br />

New patient value: Based on IBW<br />

Range:<br />

OFF or<br />

> low exhaled spontaneous tidal volume limit<br />

> low exhaled mandatory tidal volume limit<br />

and<br />

Neonatal: 5 mL to 500 mL<br />

Pediatric: 25 mL to 1500 mL<br />

Adult: 25 mL to 3000 mL<br />

New patient value: Based on IBW<br />

Resolution:<br />

1 mL <strong>for</strong> 5 mL to 100 mL<br />

5 mL <strong>for</strong> 100 mL to 400 mL<br />

10 mL <strong>for</strong> 400 mL to 3000 mL<br />

Range:<br />

OFF or<br />

Neonatal: 10 /min to 170 /min<br />

Pediatric/Adult: 10 /min to 110 /min<br />

New patient value: OFF<br />

Resolution: 1 /min<br />

16 10070808 Rev. C Software Enhancements